Test hand
Attach sample draw kit to instrument. Test for
aspirated
leakage by holding finger over end of probe.
sample
Then squeeze bulb. Bulb should stay deflated
until finger is removed.
draw kit
Use hand
Allow one squeeze of bulb for every one foot of
aspirated
hose for sample to first reach sensors. Continue
sample
to squeeze bulb once per second for 1 additional
minute to complete sampling.
draw kit
Test
Slip pump onto instrument with tubing and probe
continuous
attached. PHD6 will prompt for leak test. Hold
sample
finger over end of probe. Pump should shut down.
pump
Remove finger and press MODE to reset pump.
Use
Pump must be tested prior to use. Allow one
second for each foot of hose plus 1 additional
continuous
minute to sample the environment.
pump