3.2.2
Sterilization Instructions
Steam sterilization is safe and effective and has no contraindications for
sterilizing this equipment.
1. Individually wrap handpiece and accessories.
2. Follow the recommended minimum sterilization exposure times listed
below.
3. Use the recommended tray parameters in its instruction manual when
sterilizing handpieces and attachments in a system sterilization tray.
Handpieces and attachments may be processed in a pre-vacuum steam sterilizer
(Steam Pre-vacuum) or in a gravity (downward) displacement sterilizer (Steam
Gravity).
Minimum recommended sterilization exposure times are as follows:
Method
Steam (wrapped)
Table 1: Sterilization Parameters
Cycle
Temperature
Pre-Vacuum
270°F (132°C)
Minimum
Minimum
Exposure
4 minutes
17
Minimum
Dry Cycle
8 minutes
Need help?
Do you have a question about the D4240 and is the answer not in the manual?