3.2

Sterilization Information

3.2.1

Warnings, Precautions and Notes

1. The use of disinfecting solutions for an exterior instrument wipe
will not sterilize the equipment and is not recommended.
2. Do not sterilize equipment or accessories using Ethylene Oxide (EtO).
3. Do not sterilize equipment or accessories using a STERIS system or by
comparable sterilization methods.
4. Do not sterilize equipment or accessories in cold sterilants like CIDEX.
5. Always detach accessories from equipment prior to sterilization.
6. Do not use equipment while warm. Allow adequate time for cooling prior
to use. Cool by exposure to room temperature. Operation of equipment
that is not completely cool or dry may decrease performance and/or
reliability.
7. When sterilizing equipment and attachments in a system sterilization tray,
use the recommended sterilization parameters in its instruction manual.
8. Do not "Peel Pack" handpieces for sterilization. Sterilization in a sealed
pouch traps moisture which can cause damage.
9. The suction control valve, if one exists, must be in the fully open position
during sterilization.
10. Attachments with collet mechanisms must be sterilized with the collet
fully open.
11. An eight (8) minute minimum dry cycle must be run on all equipment and
attachments every time the product is sterilized. Failure to use a dry cycle
on the products may lead to reduced product performance or premature
product failure.
Refer also to section "1.5 Warnings and Precautions".