11.4.4 Calibrator Status
Error Queue
If an error occurs due to invalid input or buffer overflow, its error code is sent to the error
queue. The error code can be read from the queue with the command FAULT?. The error
queue holds 5 error codes. When it is empty, FAULT? returns 0. The error queue is
cleared when power is reset or when the clear command *CLS is entered.
Input Buffer
Calibrator stores all received data in the input buffer. The buffer holds 250 characters. The
characters are processed on a first in, first out basis.

11.5 Remote Commands and Error Codes

The following tables list all commands, and their descriptions, that are accepted by the calibrator.
Table 5: Common Commands
Command
*CLS
*IDN?
*RST
Table 6: Calibrator Commands
Command
CPRT_COEFA
CPRT_COEFA?
CPRT_COEFB
CPRT_COEFB?
CPRT_COEFC
CPRT_COEFC?
CPRT_R0
CPRT_R0?
DAMP
DAMP?
DISPLAY
DISPLAY?
FAULT?
ERROR_LOOP
ERROR_LOOP?
ERROR_MODE
ERROR_MODE?
ERROR_PORT
ERROR_PORT?
Description
(Clear status.) Clears the error queue
Identification query. Returns the manufacturer, model number, and firmware
revision level of the Calibrator
Resets the calibrator to the power up state
Description
Sets the custom RTD coefficient A
Returns the custom RTD coefficient A
Sets the custom RTD coefficient B
Returns the custom RTD coefficient B
Sets the custom RTD coefficient C
Returns the custom RTD coefficient C
Sets the custom RTD R0 resistance
Returns the custom RTD R0 resistance
Turns Damp on or off.
Returns if DAMP is on/off
Turns on/off the displays specified in the command
Returns which displays are on/off
Returns the most recent error code
Turns loop power on or off in percent error mode
Returns the current state of loop power in error mode
Turns percent error mode on or off
Returns whether percent error mode is on or off
Set the pressure port for percent error mode
Returns the pressure port for percent error mode
55