AirSep SeQual Regalia Instruction Manual page 23

Psa oxygen concentrator system
Table of Contents

Advertisement

Problem
No oxygen / flow
(see alarm indications
in Display of Operating
Conditions table)
Problem
Low oxygen / flow
(see alarm indications
in Display of Operating
Conditions table)
Yellow Light On
Flow Rate not visible on Flow
Indicator Display
Audible Alarm
Red Light On
Flow Indicator Blank
Back Light On
SeQual Regalia Instruction Manual
Probable Cause
Tubing blocked or kinked
Air Inlet Filter blocked
Internal component
malfunction
Probable Cause
Restriction in tubing.
Air Inlet Filter restricted
Internal component worn or
malfunctioning
Button is stuck
LCD malfunction
System malfunction
Table 6.4:
Concentrator is Running
AirSep Corporation
Solution
Clean or replace tubing.
Remove tube kink.
Clean air inlet filter.
Service required.
Solution
Repair or replace as
required.
Clean air inlet filter.
Service required.
Service required.
Service required.
Remove power cord from
wall outlet.
Service required.
23