Download Print this page

Advertisement

Quick Links

Philips IntelliVue Arm
I ns ta ll at I o n I ns t ruc t I o ns
P502673 Rev B

Advertisement

loading

Summary of Contents for Philips IntelliVue Arm

  • Page 1 Philips IntelliVue Arm I ns ta ll at I o n I ns t ruc t I o ns P502673 Rev B...
  • Page 2 IntelliVue Arm Components List Wall Track Options Select a wall track. Choose a size that is compatible with your application. Wall tracks are available in different lengths. Each has 3/8" diameter mounting holes positioned 6" center to center apart. 8326-13-XXX - 13" Wall Track 8326-19-XXX - 19"...
  • Page 3 IntelliVue Arm Components List Monitor Arm 6958-1000-XXX This arm is used for IntelliVue only. 6958-1500-XXX The minimum weight that can be placed on the arm is 32 lbs with a maximum weight capability of 53 lbs. Setup includes 1/4" Allen wrench (P701127) and 3/32"...
  • Page 4 IntelliVue Arm Assembly Instructions...
  • Page 5 IntelliVue Arm Assembly Instructions Method a step 1 Preparing wall for track mounting. The four anchorage methods illustrated in the next few pages represent load tests made to comply with California OSHPD requirements for the anchorage and installation of instrument support system.
  • Page 6 IntelliVue Arm Assembly Instructions Method B step 1 Preparing wall for track mounting. The four anchorage methods illustrated in the next few pages represent load tests made to comply with California OSHPD requirements for the anchorage and installation of instrument support system.
  • Page 7 IntelliVue Arm Assembly Instructions Method c step 1 Preparing wall for track mounting. The four anchorage methods illustrated in the next few pages represent load tests made to comply with California OSHPD requirements for the anchorage and installation of instrument support system.
  • Page 8 IntelliVue Arm Assembly Instructions Method D step 1 Preparing wall for track mounting. The four anchorage methods illustrated in the next few pages represent load tests made to comply with California OSHPD requirements for the anchorage and installation of instrument support system.
  • Page 9 IntelliVue Arm Assembly Instructions step 2 attach wall track. All tracks have 3/8" diameter mounting holes positioned 6" center to center apart. Wall Track Determine where you will be placing the Part No. 8326-#-XXX track and drill appropriate holes. Position the mount track (8326-#-XXX) on the wall.
  • Page 10 IntelliVue Arm Assembly Instructions step 3 attach and secure mount to wall track. Align the mount (8328-XXX) in the grooves at the top of the wall track. Slide it down to the desired height and tighten the set screws to lock into place using Track Mount the 3/32"...
  • Page 11 IntelliVue Arm Assembly Instructions step 4 Insert monitor arm into track mount. Insert the monitor arm (6958-#-XXX) into the track mount (8238-XXX). Monitor Arm Part No. 6958-#-XXX Track Mount Part No. 8328-XXX...
  • Page 12 (included with the IntelliVue) over the Pole adapter (61302-#-XXX) and tighten using Part No. 61302-#-XXX the (4) 1/4-20 x 1/2" Philips Flat Head Screws (P700987). Slide the collar (8240-XXX) up the pole and tighten the set screw to secure. Slide the washer (P700566) and retaining ring (P700567) onto the pole.
  • Page 13 IntelliVue Arm Assembly Instructions step 6 attach cable management to arm. Included you will receive (2) pieces of 11" cable management (P401136). Remove the backing and attach the first piece to the underside of the forearm. This will keep your cords organized. Be sure to leave enough slack for cables to loop around the elbow of the arm.
  • Page 14 IntelliVue Arm Assembly Instructions step 7 Wall Track End Cap attach cable covers to track. Part No. P104596 Included you will receive (2) wall track end caps (P104596), (1) wall track cable Wall Track Cable Cap cap (P104985) and (1) wall mount cable Part No.
  • Page 15: Height Adjustment

    IntelliVue Arm Operating Manual operating Manual Note: the monitor must be attached to the arm to make any adjustments. HEIGHt aDJustMEnt Hold the arm in a horizontal position. Use the 1/4" Allen wrench (P700146) to strengthen or weaken. Stronger Weaker JoInt FrIctIon Use the 3/32"...