Model 3000TA
Auto Mode Zero Calibration
Press Zero to enter the zero function mode.
•
Use the ▲▼ arrow keys to toggle between
•
AUTO
appears blinking.
Press Enter to begin zeroing.
•
The zeroing process will automatically conclude when the output is within the acceptable
range for a good zero. Then the analyzer automatically returns to the Analyze mode.
Manual Mode Zeroing
Press Zero to enter the Zero function.
•
Use the ▲▼ keys to toggle between
•
appears, blinking, on the display.
Press Enter to begin the zero calibration.
•
Once zero settling is completed, the instrument automatically returns to the Analyze mode.
Zero: Settling: AUTO
<ENT> To Begin
####
PPM Zero
Slope=####
ppm/s
.
.
.
####
PPM Zero
4 Left=###
ppm/s
AUTO
Zero: Settling: Man
<ENT> To Begin
####
ppm Zero
Slope=#### ppm/s
Teledyne Analytical Instruments
and
zero settling. Stop when
AUTO
MAN
and
zero settling. Stop when
MAN
Quickstart Guide
MAN
15