2.4.2 Digital Status Outputs

The M460H has four digital status outputs for indicating error and operational status
conditions. These outputs are in the form of opto-isolated open-collector transistors.
They can be used to drive status LED's on a display panel or interface to a digital device
such as a Programmable Logic Controller (PLC).
Figure 2-4 shows the most common way of connecting the digital outputs to an external
device such as PLC. Note: Most devices, such as PLC's, have internal provision for
limiting the current that the input will draw from an external device. When connecting to
a unit that does not have this feature, external dropping resistors must be used to limit the
current through the transistor output to 50mA or less.
See Chapter 7 for details on using the Status Outputs for diagnosing sensor and system-
level malfunctions.
Table 2-1 below describes the function of the status outputs.
Label
STATUS OUT 1
STATUS OUT 2
STATUS OUT 3
STATUS OUT 4
STATUS OUT 5
STATUS OUT 6
STATUS COM
Opto-Isolator
P/N 03662D Teledyne API Model 460H O 3 Monitor Operator Manual - Page 15

Table 2-1 Digital Status Output Pins

Name
Sensor O.K.
Invalid Reading
Lamp Low
Cell Dirty
Auto-Zero Error
Spare
Common Pin for
all Status Outputs
M454
Digital Output #1-4 (Collector)
Digital Output Common (Emmiter)

Figure 2-4 Digital Output Connections

Operation
Normally On
Normally Off
Normally Off
Normally Off
Normally Off
Undefined
N/A
Programmable Logic
Controller or other device
+5V
Digital
Input
Ground Provided
by PLC