GE eBike III basic Series Service Manual page 74

Electric bike with and without nibp
Table of Contents

Advertisement

Set the calibrator unit to 110 rpm.
Press NIBP (START) to start calibration step #2
Wait approx. 20 seconds, until the load display is
stable
By using the arrow keys, input the actual value of the
calibrator (e.g. 50,8 Watt) at the ergometer.
Store the value with SET.
The eBike III automatically sets the load to 400 Watt
Wait approx. 20 seconds, until the load display is
stable
By using the arrow keys, input the actual value of the
calibrator (e.g. 403,2 Watt) at the ergometer.
Store the value with SET.
The successfull completion of the dynamic load calibration
is confirmed.
Stop the calibration unit.
Press NIBP to exit the Calibration programm
Dismount the calibration unit and reassemble the cranks /
pedals to the eBike.
- 74 -
Dynamic Load Cal.
Dyn. Calibration
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
At 110 /50 W
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
Dyn. Calibration
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
At 110 /400 W
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
Calibration successfull
d
ynamic load calibration SucceSSfull
eBike III basic / comfort
1,00
3219 / 3219
110 B
0 P
50 SW
AW
START
2 -
Start
waiting for
50,8 W
1,00
3219 / 3219
110 B
0 P
50 SW
AW
SET
2 -
input load value
1,00
3219 / 3219
110 B
0 P
400 SW
AW
START
3 -
Start
waiting for
403,2 W
1,00
3219 / 3219
110 B
0 P
400 SW
AW
SET
3 -
input load value
EXIT
2018111-336 Rev. B

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Ebike iii comfort series