Set the calibrator unit to 110 rpm.
•
Press NIBP (START) to start calibration step #2
•
Wait approx. 20 seconds, until the load display is
stable
•
By using the arrow keys, input the actual value of the
calibrator (e.g. 50,8 Watt) at the ergometer.
•
Store the value with SET.
•
The eBike III automatically sets the load to 400 Watt
•
Wait approx. 20 seconds, until the load display is
stable
•
By using the arrow keys, input the actual value of the
calibrator (e.g. 403,2 Watt) at the ergometer.
•
Store the value with SET.
The successfull completion of the dynamic load calibration
is confirmed.
•
Stop the calibration unit.
•
Press NIBP to exit the Calibration programm
Dismount the calibration unit and reassemble the cranks /
pedals to the eBike.
- 74 -
Dynamic Load Cal.
Dyn. Calibration
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
At 110 /50 W
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
Dyn. Calibration
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
At 110 /400 W
Gain:
DMS:
RPM:
Load:
d
ynamic load calibration Step
Dynamic Load Cal.
Calibration successfull
d
ynamic load calibration SucceSSfull
eBike III basic / comfort
1,00
3219 / 3219
110 B
0 P
50 SW
AW
START
2 -
Start
waiting for
50,8 W
1,00
3219 / 3219
110 B
0 P
50 SW
AW
SET
2 -
input load value
1,00
3219 / 3219
110 B
0 P
400 SW
AW
START
3 -
Start
waiting for
403,2 W
1,00
3219 / 3219
110 B
0 P
400 SW
AW
SET
3 -
input load value
EXIT