Dynamic Load Cal - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

d
l
C
.
ynaMIC
oad
al
To perform a dynamic load calibration, a calibrated
dynamic test bench is needed.
The dynamic calibration of the load unit is described in the
session "„Dynamic Load Calibration" on page 72.
s
u
oftware
pdate
For a description of a software update of the ergometer
(firmware) see the session „Software update" on page
78.
B
8n2
audrate
The baudrate can be selected between 1200 Baud and
115000 Baud.
Default value for standard ergometers is 4800 Baud.
Select the desired parameter using the arrow keys and
confirm by pressing NIBP.
s
C
addle
alIBratIon
(eBike III comfort only)
This parameter is used to adjust the electrical saddle motor
(display scale 1-40).
Select "SET TOP Position" and confirm by pressing NIBP.
Use the arrow keys to move the saddle to the desired top
position (99 cm ± 0,5 cm above bottom plate; see picture)
and confirm "SET" by pessing NIBP.
- 64 -
Baudrate
S
election of Serial baudrate
Saddle Calibration
SET TOP Position
SET BOTTOM Position
EXIT
Saddle Calibration
TOP Position
Adjust with  
SET
c
alibration of Saddle height
eBike III basic / comfort
8N2
1.200
2.400
4.800
9.600
19.200
38.400
76.800
Height: 22
top p
oSition
2018111-336 Rev. B