IMMUNITY test
Electrostatic discharge
(ESD) IEC 61000-4-2
Electrical fast transient/
burst IEC 61000-4-4
SurgeIEC 61000-4-5
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
Power
frequency(50/60 Hz)
magnetic fieldIEC
61000-4-8
Note: UT is the a.c. mains voltage prior to application of the test level.
Declaration – electromagnetic immunity
The Philips Avent breast pump is intended for use in the electromagnetic environment
specified below. The user of the Philips Avent breast pump should assure that it is used in
such an environment.
IEC 60601
Compliance
test level
level
± 6 kV contact
± 6 kV contact
± 8 kV air
± 8 kV air
± 2 kV for
± 2 kV for
power supply
power supply
lines
lines
± 1 kV line(s) to
± 1 kV line(s)
line(s)
to line(s)
<5 % UT (>95
<5 % UT (>95
% dip in UT)
% dip in UT)
for 0,5 cycle40
for 0,5 cycle70
% UT (60 %
% UT (30 % dip
dip in UT) for
in UT) for 25
5 cycles70 %
cycles<5 % UT
UT (30 % dip
(>95 % dip in
in UT) for 25
UT) for 5 s
cycles<5 % UT
(>95 % dip in
UT) for 5 s
3 A/m
3 A/m
Electromagnetic
environment – guidance
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be
at least 30 %.
Mains power quality should be
that of a typical residential or
hospital environment.
Mains power quality should be
that of a typical residential or
hospital environment.
Mains power quality should
be that of a typical residential
or hospital environment. If
the user of the Philips Avent
breast pump requires continued
operation during power mains
interruptions, it is recommended
that the Philips Avent breast
pump be powered from an
uninterruptible power supply
or a battery.Interruptible
electric supply can cause the
breast pump to turn off. This is
acceptable as this will not lead
to unacceptable risks.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a
typical residential or hospital
environment.