Miscellaneous Parts; Adjustments - HP 85325A Operating And Service Manual

R/q/u/v/w millimeter subsystem, millimeter-wave interface kit
Table of Contents

Advertisement

Miscellaneous Parts

Table 6-6
Service
Replaceable Parts
The following parts are supplied with each system.
Miscellaneous Parts
Description
Screws, captive
Screws, 1/2 inch
Screws, 3/8 inch
Nut, hex 4-40
Nut, hex 4-40
RF cable, SMA (m) / (m), 1 m
Wrench, 1/4 inch
Wrench, 3/16 inch
Torx driver T10
2
Modified wrench, 5/16
Attenuator, fixed, 10 dB, 3.5 mm
Operating and Service Manual
1. Only supplied with R band.
2. This wrench is specially designed. The "closed" end has been
machined to be smaller, and has been slotted so it will fit over a cable.
The modified end has another feature as well, it slips when
approximately 10 inch-pounds of force is exerted on the SMA
connection. 10 inch-pounds is the torque specification for all SMA
connections inside the mm-wave modules, or inside the HP 85309A.

Adjustments

Only the HP 85309A LO/IF unit requires periodic circuitry adjustments.
Refer to the HP 85310A or HP 85301B Operating and Service Manual for
HP 85309A adjustment procedures. If the LO and RF source's timebases are
not connected together, refer to either of the manuals (mentioned above) for
adjustment instructions.
The LO power to the test and reference modules must be adjusted during
installation. Refer to Installation, Chapter 1 for more instructions.
6-14 HP 85325A Operating and Service Manual
HP Part Number
1390-0765
3030-0209
3030-0221
2260-0002
1
2260-0009
5061-5458
8710-0510
8720-0013
8710-1623
08555-20097
HP 8493C, opt. 010
85325-90001