Technical Data - Jct JES-370 Operating Manual

The heated gas sampling probe
Table of Contents

Advertisement

Manual JES-370
––––––––––––––––––––––––––
BA_DE_JES370_v1.3
Flow rate (back purging)
Air for back purge
Protection class
Area classification
Dimension over all
Junction box
Dead volume
Mounting flange *1
Option
Mounting angle
Mounting position
Weight
housing material, colour
ABS filter
Sample gas inlet
Sample gas outlet
calibration port
(optional)
connection condensate
drain (option)
Back purge port
Back purge gas wetted ma-
terials
Required amount of instru-
ment air
Temperature sensor
heater element
Power consumption
Freeze protection
Valve back purge
Pilot valve
Electrical actuator
power consumption
power supply
electrical connection
Power supply
Electrical connection
Switching alarm relay of
back purge controller
Approval / Sign
––––––––––––––––––––––––––
[ 10 / 30 ]
approx. 90 Nm³/h @ 4 bar
160 Nm³/h @ 7 bar
instrument air according to
ISO 8573-1 class 1.2.1
Electronic: IP 65
Probe: IP 43
For use in safe, non hazar-
dous area only
Construction
696x613x443 mm WxhxD
340x150x100 mm hxWxD
177 cm
3
DN 65, PN 6, form A accor-
ding to EN 1092-1; SS316Ti;
threaded bolts M12x41 mm
2"ANSI; 150lbs., hole pattern
according to ASME B16.5;
threaded bolts M16x48 mm
range 5° to 10° with respect to
the horizontal, sloping down
torsion angle max. 10°
approx. 30 kg
SS304, stainless steel natural
Filter balls of borosilcate
glass
G3/4" female thread
PVDF hose fitting DN 4/6
mm
6 mm bulkhead union,
SS316
PVDF hose fitting DN 4/6
mm
12 mm bulkhead union,
SS316
SS316, brass, EPDM
max. 1500 l/min respectively
max. 450 l/cycle @ 7 bara
inlet pressure
Pt100
800 W
approx . 810 VA
125 - 250 W
9 W, 24 or 115 or 230 VAc
+/- 10%
5 W, 24 or 115 or 230 VAc
+/- 10%
6 VA
230 VAc
clamps max. 1,5 mm
230 VAc/50 hz +/- 10% or
115 VAc/60 hz +/- 10%
Springtype terminal
clamping range 0,08...2,5 mm
Typ. 230 VAc/2 A/
min. 5 VADc/5 mA
cE
2
2

Advertisement

Table of Contents
loading