Burkert 8220 Operating Instructions Manual

Burkert 8220 Operating Instructions Manual

Conductivity sensor

Advertisement

Quick Links

Type 8220
Leitfähigkeitssensor
Conductivity sensor
Capteur de conductivité
Operating Instructions
Bedienungsanleitung
Manuel d'utilisation

Advertisement

Table of Contents
loading

Summary of Contents for Burkert 8220

  • Page 1 Type 8220 Leitfähigkeitssensor Conductivity sensor Capteur de conductivité Operating Instructions Bedienungsanleitung Manuel d'utilisation...
  • Page 2 We reserve the right to make technical changes without notice. Technische Änderungen vorbehalten. Sous réserve de modifications techniques. © Bürkert SAS, 2014 - 2018 Operating Instructions 1809/01_EU-ML 00273224 / Original FR...
  • Page 3: Table Of Contents

    Type 8220  1. About the operAting instructions ........4 6.7. sensor data ..................11 1.1. Definition of the word "device" ............4 7. instAllAtion AnD Wiring ..............11 1.2. symbols used ..................4 7.1. safety instructions................11 2. intenDeD use ....................5 7.2. installation on the pipe ..............12 7.3. electrical wiring ................. 13 3. bAsic sAfety informAtion ...............5 7.3.1. Safety instructions ..............13 4. generAl informAtion ................7...
  • Page 4: About The Operating Instructions

    ▶ The Operating Instructions must be read and understood. Warns against material damage. 1.1 Definition of the word "device" Advice or important recommendations. The word "device" used in the Operating Instructions always refers to the conductivity sensor type 8220. refers to information contained in the Operating 1.2 symbols used Instructions or in other documents. danger →...
  • Page 5: Intended Use

    Type 8220 Intendeduse inTenDeD use bAsic sAfeTy informATion This safety information does not take into account: use of the device that does not comply with the • any contingencies or occurences that may arise during instal- instructions could present risks to people, nearby lation, use and maintenance. installations and the environment. • the local safety regulations for which the operating company the 8220 conductivity sensor is intended solely for the is responsible including the staff in charge of installation.
  • Page 6 Type 8220 Basicsafetyinformation nOTe elements / components sensitive to electrostatic discharges ▶ Do not use this device in an explosive atmosphere. ▶ This device contains electronic components sensitive to ▶ Do not use this device in an environment incompatible electrostatic discharges. They may be damaged if they are with the materials it is made of.
  • Page 7: General Information

    The device is made up of: F-67220 TRIEMBACH-AU-VAL • a measuring sensor of physical quantities, made up of: You may also contact your local Burkert sales office. - 2 electrodes which measure an impedance in Ohm; The addresses of our international sales offices are available on the internet at: www.burkert.com...
  • Page 8: Description Of The Name Plate

    Type 8220 Technicaldata 5.3 Description of the name plate TechnicAl DATA 6.1 conditions of use COND :8220-FKM-PVDF Ambient temperature 0 to +60°C ≤ 80 %, non-condensing Air humidity S/N 5721 00426874 W48ML 6.2 conformity to standards and directives The applied standards, which verify conformity with the EU Fig. 1: Name plate of the device (example) directives, can be found on the EU-type examination certif- icate and/or the EU declaration of conformity (if applicable).
  • Page 9: Materials

    Type 8220 Technicaldata • Device used on a piping (PS = maximum admissible pressure; 6.4 materials DN = nominal diameter of the pipe) element material type of fluid conditions Housing Screw of the connector stainless steel Fluid group 1, Article 4, Paragraph 1.c.i DN ≤ 25...
  • Page 10: Fluid Data, Measurement Data

    Type 8220 Technicaldata 6.6 fluid data, measurement data Pressure (bar) Pipe diameter DN15 to DN200 metal Process connection S020 fitting for analysis purposes PVDF Fluid pressure PN 10 PVDF (PN10) Fluid temperature and material of PVC + PP the fitting S020 used, can limit the max.
  • Page 11: Sensor Data

    Type 8220 Installationandwiring 6.7 sensor data insTAllATion AnD Wiring Sensor K=0,01 7.1 safety instructions • Measurement range • 0,05 to 20 µS/cm danger • Type of fluid • ultra pure water, pure Danger due to electrical voltage water ▶ Shut down the electrical power source of all the conduc- Sensor K=0,1 tors and isolate it before carrying out work on the system.
  • Page 12: Installation On The Pipe

    Type 8220 Installationandwiring 7.2 installation on the pipe Warning The device can be installed on a pipe using a fitting type risk of injury due to non-conforming installation. S020 (see Fig. 3 et Fig. 4). ▶ The electrical and fluid installation can only be carried out by qualified and skilled staff with the appropriate tools.
  • Page 13: Electrical Wiring

    If this cannot be avoided, observe → Slowly push the device in the fitting. If a min. distance of 30 cm. the mounting is correctly done the device cannot be turned around anymore. → Hand thighten the nut to lock the assembly. Fig. 4: Installation of the 8220 on the S020 fitting English...
  • Page 14: Technical Data Of The Cables For The Female Connectors, Type 2508 Or Type 2509

    Type 8220 Installationandwiring 7.3.2 Technical data of the cables for the 7.3.3 Assembling the female connector female connectors, type 2508 or → Loosen the nut [1] of the cable gland. type 2509 → Remove the terminal block [3] from the cable technical data recommended housing [2]. Electromagnetic protection shielded (EMC) → Insert the cable in the nut [1], in the seal [4],...
  • Page 15: Ensuring The Equipotentiality Of The Installation

    Type 8220 Installationandwiring 7.3.4 ensuring the equipotentiality of the Electrical power supply installation To ensure the equipotentiality of the installation (power supply Transmitter - device - fluid): PE/FE → Connect the different earth points to one another of the installation to eliminate the potential differences that may 8220 occur between themselves.
  • Page 16: Connect The Device To The Transmitter

    Terminal block denser between the negative power supply terminal and the earth. of the connector Fig. 7: Equipotentiality of the installation with plastic pipes 2508 1 2 3 4 5 6 7 8 9 Conductivity module of the FE = functional earth transmitter 8619 terminal Fig. 9: Connection example of the 8220 to the transmitter 8619 English...
  • Page 17: Commissioning

    Type 8220 Commissioning commissioning mAinTenAnce AnD TroubleshooTing Warning risk of injury due to non-conforming commissioning. 9.1 safety instructions Non-conforming commissioning could lead to injuries and danger damage the device and its surroundings. Danger due to electrical voltage ▶ The staff in charge of commissioning must have read and understood the contents of the Operating Instructions.
  • Page 18: Maintenance Of The Device And The Conductivity Sensor

    Type 8220 Sparepartsandaccessories spAre pArTs AnD Warning Accessories risk of injury due to non-conforming maintenance. CaUTiOn ▶ Maintenance must only be carried out by qualified and skilled staff with the appropriate tools. risk of injury and/or material damage caused by the ▶ Ensure that the restart of the installation is controlled after use of unsuitable parts. any interventions. Incorrect accessories and unsuitable replacement parts may cause injuries and damage the device and the sur- 9.2...
  • Page 19: Transport, Storage, Disposal

    Type 8220 Transport,storage,disposal TrAnsporT, sTorAge, DisposAl nOTe Damage due to transport NPT 1/2 Transport may damage an insufficiently protected device. • Transport the device in shock-resistant packaging and away from humidity and dirt. • Do not expose the device to temperatures that may exceed the admissible storage temperature range.
  • Page 20 Type 8220  English...
  • Page 22 www.burkert.com...