Maintenance

6
Maintenance
Dräger SAM 3100/3200 aspirator system:
The aspirator system is designed so as to minimise the need
for servicing.
Directions for maintenance and servicing can be found in
the Instructions for Use for the components used.
Depending on the dust load in the sample gas, the filter
inserts should be replaced when visibly soiled. Spare parts:
see the Instructions for Use for the filter.
In the case of systems with gas sampling pump, it is
recommended that the diaphragm and valve plates of the
pump are replaced at least once a year.
Spare parts: refer to the Instructions for Use for the gas
sampling pump.
The ejectors for delivering sample gas have no mechanical
parts subject to wear, and are therefore maintenance-free.
Gas transmitter
Maintenance and servicing of the gas transmitter should be
carried out according to the enclosed Instructions for Use.
The calibration interval and calibration instructions can be
found in the Instructions for Use for the gas transmitter and
for the DrägerSensor.
For aspirator systems with gas sampling pump, the pump
should be switched off during test gas feeding (calibrating).
For models with ejector, the air feed should likewise
be interrupted.
7

Wiring diagrams

CAUTION
!
Electrical wiring is only to be laid and connected by an
expert, paying attention to the relevant regulations and
laws concerning electrical equipment in potentially
explosive atmospheres as well as the approval
conditions.
Consult the Instructions for Use for installation!