User Guide
3
Electrical Connections
3.1 General Connections
The electrical connections for the RT8A thermometer power supply and
temperature output are made via the 6-way socket on the rear of the
thermometer (see Fig. 6).
The connection to the thermometer can be made via any of the following:
•
Pre-wired 4m cable (Land Part No. 029.673)
•
6-way plug (Land Part No. 206.551)
•
6-way plug housed in protective jacket end cap (Land Part No. 091.562)
Thermometer
Pin
1
2
3
4
5
6
To connect the 6-way plug to the RT8A thermometer (see Fig. 4):
•
Align the red marker near the lugs of the plug, with the red marker
located above the keyway in the thermometer socket.
•
Push the plug into the socket, ensuring that the locking sleeve slides
forwards, locking the plug to the socket.
To disconnect the 6-way plug to the RT8A thermometer:
•
Grip the locking sleeve portion of the plug.
•
Slide the locking sleeve rearwards to release the locking mechanism and
disconnect the plug from the thermometer socket.
RT8A
Function
-
-
V+, 4 to 20mA
-
-
V-, 4 to 20mA
Cable Colour
End Cap Pin No.
-
-
Red
-
-
Black
Thermometer
-
-
5
-
-
6
Page 5