5.2 Services

When any cooling or purging accessories are utilised, it is important that the
air and water supply to these facilities are fully conditioned and supplied at
the recommended flow rates. The following information details the specified
flow rates:
S4-J Cooling Jacket Assembly
Requirement: Conditioned water supply (clean and free from any
contaminants) Flow rate: 1ltr/min or 0.035cu.ft/min
S4-P Air Purge Assembly
Requirement: Conditioned air supply (clean and dry, free from any
contaminants) Flow rate: 1ltr/sec or 2cu.ft/min
CAUTION
"Over-cooling" of the thermometer may cause a build-up of condensation
on or inside the instrument.
To prevent this, ensure that the water supply temperature is not lower
than the local dewpoint temperature.
Page 12