  • Page 1 20000005398 A11 basic_052018 ® A11 basic...
  • Page 2: Table Of Contents

    Always wear your personal protective equipment (protective gloves, etc.). Please note that wear debris from the appliance or rotating accessories can end up in the medium to be processed. Please contact IKA ® should you have any questions...
  • Page 3: Correct Use

    The analytical mill A 11 basic may not been used in explosive atmospheres or well, because of the closed design, the loss of ground material is equivalent to under water. Do not crush explosive, toxic or noxious substances. Please observe practically zero.
  • Page 4 Please note: Materials that are not listed above should only be reduced in the If the feed product has a fine fraction or if the same sample is cooled several A 11 basic after confirmation from IKA ® , especially if there is uncertainty about the times, use filling funnel A 11.7 (IKA...
  • Page 5: Commissioning

    Caution: Liquid nitrogen cooling is not permitted in grinding beaker A 11.4 (250 thumb (figure 2), and the mill starts. Pressing the drive down softly and slowly in ml beaker made of polycarbonate).The beaker would break during cooling. the axial pressing direction forces the pulverization tool into the grinding material (pressing on one side only can tilt the drive, which could result in premature wear and tear on the shaft seal).
  • Page 6: Malfunctions And Eliminating Malfunctions

    If one side is dull or worn, you can use the other side by turning the pin is inserted into the membrane grommet on the back of the A11 basic (above the beater or blade around.
  • Page 7 Causes: Caution: The shaft of the grinding container and the pulverization - Too much grinding material in the grinding container (filling hight exceeded). tools may not be hot. If necessary, allow the mill to cool off. - The feed granularity is too coarse. - Motor circuit breaker is tripped because of lock (overload).
  • Page 8: Maintenance And Cleaning

    Finally the desired pulverization tool is assembled. Maitenance and cleaning Maintenance: Cleaning: If beaters or blades wear out over the course of time, especially as a result of Unplug the power cord before all cleaning tasks! constant grinding of very hard materials, so that rapid grinding with the appropriate quality is no longer possible, you need to replace the worn - out parts.
  • Page 9: Accessories

    Fuels water containing tenside The maximum filling height, shown in figure 5, must not be exceeded (about 20 mm For materials which are not listed, please request information from IKA ® . Wear the below the stainless steel upper edge). Otherwise the drive may overheat.
  • Page 10: Warranty

    Orange-red glow lights, display of over- Warranty temperature and grinding area not closed Drive Universal motor You have purchased an original IKA ® laboratory machine which meets the highest engineering and quality standards. Parts in contact with product In accordance with IKA ®...