8.3 Change of filter
Attention:
 Contaminated filters have to be changed under all corresponding safety
measures
 Filter change only with a machine switched off
 Use approved filters only
 Do not use bonding agents on the machine
Remove the hepa-filter:
 Remove the prefilter
 Unscrew the hexagone screws M8 and remove the tension frame
 Remove the hepafilter and dispose of it professionally
Insertion of the hepa-filter:
 Control the sealing face of the machine and clean.
 Clean the inside of the housingI
 Insert the new filter in the middle
 Tighten the tension frame and the hexagone screws M8 (tighten the screws carefully)
 Insert the pre-filter
Important:
Treat the hepafilter with care because otherwise damages can suspend the
effectivity of the filtration.