Quick Start Guide
14. Wait while the gas concentration measurement stabilizes.
Refer to
Figure 6-2: Typical Calibration Profile
15. Select Next when the gas concentration measurement stabilizes at
or near the target gas concentration level.
16. Wait while the AMS Wireless Configurator calibrates.
When the calibration process finishes, the new adjusted reading is
displayed.
17. Select Next.
18. Select Accept calibration.
19. Select Next.
The Service Reminder screen is displayed if a service reminder is
configured and enabled.
20. Select Next to accept the service reminder date or enter another
date.
Refer to the Service Reminders section of the Rosemount 928 Wireless
Gas Monitor
21. Shut off the target gas flow at the regulator.
22. Detach the calibration tubing from the regulator on the target gas
source and from the IP filter inlet on the bottom of the sensor.
34
Figure
6-2.
A. Gas concentration ppm
B. Time (in seconds)
C. Gas concentration measurement has stabilized
Reference Manual
for more information.
Emerson.com/Rosemount