ALARMS 
No Breathing CO2 
High End Tidal CO2 
 
Low End Tidal CO2  
 
Alarms can be silenced by touching the screen. 
The above alarms are self‐canceling when the cause of the alarm is 
resolved. 
 
 
The CO2 capnograph waveform amplitude 
is insufficient for breaths to be detected. 
 
Possible causes include:  the ventilator is 
not functioning properly, there is a 
ventilation tubing disconnect, the ET tube is 
not in the patient or the patient is not 
breathing for whatever reason, etc, etc. 
 
Identify and correct the problem. 
Displayed when ETCO2 > 60  mmHg 
 
This can be caused by the patient not being 
adequately ventilated due to airway 
problems, ventilation tubing problems, 
ventilator settings, bagging frequency 
and/or by a problem/malfunction with the 
ventilation system including the setup, the 
valves, the CO2 absorber or other 
components. Refer to CO2 monitoring 
educational material for all causes and 
cures. 
 
Identify the problem(s) and correct. 
Displayed when ETCO2 < 15 mmHg 
 
This can be caused by the patient being 
over ventilated due to ventilator settings, 
bagging frequency and/or by a 
problem/malfunction with the ventilation 
system including the setup, the valves, the 
CO2 absorber or other components. Refer 
to CO2 monitoring educational material for 
all causes and cures. 
 
Identify the problem(s) and correct. 