Operating Instructions/User Selections.
Note 1:  The CO2 Module is shipped with the large airway adapter attached 
to the sensor.  The sensor has been "zeroed" with that airway adapter 
attached prior to shipment and should not need to be zeroed again before 
use.  However, anytime the airway adapter is changed on the sensor, the 
sensor must be "zeroed" with the new airway adapter in place. 
Zeroing the sensor: 
1. Ensure no CO2 gas is in the airway adapter. 
2. Access the CO2 sensor zero screen on the petMAP+II setup menu 
by pressing and holding your finger on bottom left quadrant of the 
touch screen while powering ON the petMAP+II with the power 
button.  Release the touch screen at the count of 10 to enter the 
CO2 sensor zeroing mode. 
3. Observe the on‐screen messages for warm‐up and zeroing. Both 
warm‐up and zeroing should complete within 2 minutes each. 
4.
Wait for the petMAP+II to report "Finished Zeroing" on the 
display—the petMAP+II will automatically power OFF.  Power the 
petMAP+II unit ON and turn on the CO2 function. Verify the 
displayed instantaneous CO2 value (not the ETCO2 CO2) is 0 or 1, 
and if not, re‐zero the device. 
   
 
3.
Attach the sensor to the 
airway adapter.  
****Note that the airway and 
sensor must be in the upright 
vertical position as shown.
 
 
   
 
12 