8 Automated cleaning cycle parameters
Note:
The washer/disinfector should fulfill the requirements as
specified in ISO 15883.
Step
Rinse
Pre-wash
Cleaning
Rinse
Thermal disinfection
Dry
9 Inspect the device.
Remove all the devices from the washing basket. Inspect
the cannulations, coupling sleeves, etc. for visible soil. If
necessary, repeat the manual pre-clean/automated cleaning
cycle. Confirm that all parts are completely dry.
Mechanical cleaning/disinfection is an additional stress
for power equipment, especially for seals and bearings.
Therefore, systems must be properly lubricated and
regularly sent to be serviced (at least once per year).
Duration
Cleaning instructions
(minimum)
2 minutes
Cold tap water
1 minute
Warm water (≥ 40 °C); use detergent
2 minutes
Warm water (≥ 45 °C); use detergent
5 minutes
Rinse with de-ionized (DI) or purified water (PURW)
5 minutes
Hot DI water, ≥ 93 °C
40 minutes
≥ 90 °C