ADInstruments Human NIBP Nano Owner's Manual page 60

Non-invasive hemodynamics
Hide thumbs Also See for Human NIBP Nano:
Table of Contents

Advertisement

Figure A–10
Finger and brachial artery
pressure waveforms in
a woman aged 73. [Data
from a study in elderly
healthy volunteers by
Rongen et al., 1995.]
Figure A–11
Finger and brachial artery
pressure waveforms in a
man aged 27. [Data from
study of Harms et al,
1999.].
55
200
150
100
50
Finger pressure
Finger pressure
0
25
26
200
150
100
50
Finger pressure
0
25
26
Pressure gradient
Mean finger blood pressure is lower than mean intra-arterial brachial
pressure due to a pressure gradient over the arm arteries caused by flow.
This pressure decay was found to increase with age, whereas pulse wave
amplification is highest in the young. Due to the combined eff ects of pulse
wave amplification and pressure decay, systolic pressure is elevated in
the young but equal to or even lower than brachial pressure in elderly
subjects. The mean pressure gradient is clearly present in Figure A–10 and
Figure A–11.
Bias and precision
In a group of 53 subjects, described in previous studies [Bos 1992, Idema
1987, Rongen 1995] the following data were obtained for brachial and
finger pressure:
Brachial pressure
Brachial pressure
27
28
29
Time (s)
Time (s)
Brachial pressure
Brachial pressure
Finger pressure
27
28
29
Time (s)
Time (s)