Electrical Requirements; Plumbing Requirements - Thermo Scientific Merlin M75 Manual

Recirculating chiller
Hide thumbs Also See for Merlin M75:
Table of Contents

Advertisement

Chapter 2
Electrical
Requirements
DANGER
Plumbing
Requirements
2 - 2
Merlin
The chiller construction provides protection against the risk of
electrical shock by grounding appropriate metal parts. The protection
will not function unless the power cord is connected to a properly
grounded outlet. It is the user's responsibility to assure a proper
ground connection is provided.
Amperage based on extreme operating conditions. Operating in a 20°C
ambient and at nominal voltage will result in a lower amperage draw.
Refer to the name plate label on the rear of the chiller for specific electrical
requirements
Chiller
Pump
M 75
PD 2
The plumbing connections are located on the rear of the chiller and are
labeled
(supply) and
Pipe Thread. Chillers are supplied with
Remove the plastic protective plugs
from both plumbing connections.
If requires, install the barbed
adapters to these connections.
Do not install the adapters or your
fittings dry. Use two wraps of
Teflon
tape on the male thread
®
wound in a clockwise direction
while facing the thread, or a suitable
Teflon
sealant/lubricant. The ½"
®
male connection can be tightened
2 full turns ±¼ turn to seal.
Overtightening will damage the
female thread.
Connect
to the hose feeding the inlet of your application. Connect
to the hose from the outlet of your application. Clamp all connections.
Voltage
Current Value
208/230
10.1
(return). The connections are ½ inch Female
/
inch and ½ inch barbed adapters.
3
8
Line Cord Plug
6-15P
Drain
Thermo Scientific