Site; Direction Of Airflow; Chapter 2 Installation And Operation - Thermo Scientific Merlin M75 Manual

Recirculating chiller
Hide thumbs Also See for Merlin M75:
Table of Contents

Advertisement

CAUTION

Direction of Airflow

Thermo Scientific
Chapter 2

Site

Ambient Temperature Range*
Relative Humidity Range
Operating Altitude
Overvoltage Category
Pollution Degree
*The chiller retains its full rated capacity in ambient temperatures up to approximately 25°C (77°F).
Reduce the cooling capacity 1% for every 0.5°C (1°F) above 25°C (77°F), up to a maximum ambient
temperature of 35°C (94°F).
Never place the chiller in a location where excessive heat, moisture, or
corrosive materials are present.
The chiller has an air-cooled refrigeration system. Air is drawn through
the front and discharged through rear and side panels. The chiller must
be positioned so the intake and discharge are not impeded. A minimum
clearance of 3 feet (1 meter) on all vented sides is necessary for adequate
ventilation. Inadequate ventilation will cause a reduction in cooling capacity
and, in extreme cases, compressor failure.
Excessively dusty areas should be avoided and a periodic cleaning schedule
should be instituted (see Chapter 4, Maintenance). Optional air filters are
available, contact our Service Department. See Preface, After-Sale Support.
Installation and Operation
10% to 80% (non-condensing)
Sea Level to 6000 feet (1830 meters)
10°C to 35°C (50°F to 94°F)
II
2
Merlin 2 - 1