Description; General - BRUEL & KJAER 4152 Instructions And Applications

Artificial ear for measurement on earphones as used in hearing aids and telephone communication systems. provided with couplers made in accordance with the leg recommendation 126, and various american and german standards.
Hide thumbs Also See for 4152:
Table of Contents

Advertisement

1.
Description
General.
The Artificial Ear Type 4152 is designed to enable
acoustical
measurements
on earphones to be carried out under well-defined acoustical
conditions.
It
consists basically of a replaceable acoustical coupler, and two sockets for
the mounting
of a Condenser Microphone
Cartridge Type 4144,
and a
Micro-
phone Preamplifier Type 2619
or
the
Sound
Level Meter Type 2203.
Fig. 1.1. The Artificial Ear Type 4152.
A spring arrangement is provided to fulfil certain Standards requirements
regarding the force applied to the object under measurement.
It
can be
adjusted to apply a force of 200 grammes to an
earphone
when using the
2 cm s Coupler, as prescribed in
the
earlier American
Standard ANSI Z.