Bante210 Benchtop pH Meter

Setup Menu

The Bante210 pH meter contains an integrated setup menu that is used to customize the displayed option to meet measurement requirements. The following
table describes the functions of the menu items.
MENU
DESCRIPTION
Set the pH buffer group for calibration and auto-recognition.
Set the number of calibration points.
Set the default temperature unit.
When the option is enabled, the meter will automatically
sense a stable reading and lock the measurements.
When the option is enabled, the meter will automatically
switch off if no key is pressed within 180 minutes.
When the option is enabled, all of the calibration data and
selected parameters will back to factory default settings,
the meter must be recalibrated.
Setting a default option
1.
In the measurement mode, press and hold the
2.
Press the  or  key select the desired option.
3.
Press the
key to confirm and move to the next menu item.
Enter
4.
Repeat the steps above until the meter returns to the measurement mode. Setting is completed.
During the setting process, press the
Menu item
key for 3 seconds to enter the setup menu.
key, the meter will exit the setting and return to the measurement mode.
Meas
OPTIONS
USA (pH4.01/7.00/10.01)
NIST (pH4.01/6.86/9.18)
1 point
2 points
3 points
Degrees Celsius
Degrees Fahrenheit
Enable
Disable
Enable
Disable
Enable
Disable
DEFAULT
USA
3 points
° C
Disable
Disable
Disable
Option