Download Print this page

Braun ExactFit 5 Connect Manual page 6

Hide thumbs Also See for ExactFit 5 Connect:

Advertisement

What to do if .....
Problem
Reason
Heart rate
Appears in the measurement
symbol
pulse is detected.
Low battery
Appears when the battery
indicator
voltage is excessively low or the
positions of the batteries are
incorrect.
Measuring
Appears when the accurate
error
blood pressure and pulse could
not be obtained.
E0 shows
The measurement process does
not detect the pulse signal.
E1 shows
The cuff is not secure.
E2 shows
The result
out of measurement range.
E3 shows
Inflation timeout when you
measurement.
E254
Calibration error or other
hardware problem.
E255
Calibrate device errors or store
read and write errors.
E256
Bluetooth upload failed
10
Solution
• Measurement in progress, remain quiet.
• Replace all four batteries with new ones.
Insert the batteries in the correct
positions. Be aware of +/- positions.
• Press "start/stop" button again and
remeasure.
to instructions.
• Check that there are no kinks in tube.
• Check that hand is relaxed.
• Check if talking or moving
during measurement.
• Check if posture is correct.
Loosen the clothing on the arm and then
measure again.
Refasten the cuff and then measure again.
• Relax for a moment and then measure again.
• Loosen the clothing on the arm and then
measure again.
• Reload the battery for re-measurement, if the
problem persistis,contact the retailer or our
customer service department for further
assistance,if you need return or repair, please
see the instructions on the warranty card.
Reload the battery for re-measurement,if the
problem persistis,contact the retailer or our
customer service department for further
assistance.
Restart the device and Re-operate according
to the update firmware operation mode.
Method of measurement
Oscillometric
Model number
BUA6350
Range of measurement
Pressure 0~299 mmHg
Pulse 40~199 beats/minute
Accuracy
Pressure +/- 3 mmHg
Pulse +/- 5 % Max.
Deluxe automatic
Display
LCD with backlight
Sets of device memory
60 sets per user
Sets of Bluetooth memory
200 sets per user
Operating temperature
+10 °C ~ +40 °C, less than 85 % R.H.
Storage temperature
-20 °C ~ +60 °C, less than 85 % R.H.
Operating atmospheric pressure
860-1060 hPa
Unit weight
Approximately 500gr (without batteries)
Power supply
Alkaline battery: 4 x AA (LR6) 1.5V
Battery life
300 times measurement
Whenever not used for 1 minute
Accessories
quick reference guide, storage pouch
Wireless communication
Bluetooth® Smart
Frequency Range: 2.4GHz ISM Band (2400 - 2483.5 MHz)
Modulation: GFSK
Service life:
5 years.
Important
Read the operating instructions.
pressure ranges, the technical accuracy of the measurement cannot be guaranteed.
Internally powered equipment
Type BF applied parts
Protected against solid foreign objects of 12.5 mm diameter and larger.
Protected against water drops falling vertically when the device is tilted up to 15°.
oxide.
Continuous operation with short-time loading.
40 °C
Operating temperature
10 °C
Storage humidity
Subject to change without notice.