Testing a Quality Control Solution
Each new operator
Before using the StatSensor Meter for the first time
When opening a new vial of test strips
If a patient test has been repeated and the blood
creatinine results are still lower or higher than
expected
After cleaning the test strip port on the meter
If there are other indications that the system is not
working properly
Whenever problems (storage, operator, instrument)
are identified or anytime there is a concern the accu-
racy of the meter may have been affected by rough
handling (such as dropping the meter).
As required by the institution's quality control policy
or local regulatory requirements
34