This document explains how to connect to the NI 9215. In this document, the NI 9215 with screw terminal, NI 9215 with spring terminal, and NI 9215 with BNC are referred to inclusively as the NI 9215. Before you begin, complete the software and...
Page 3
If the product is damaged, return it to NI for repair. NI 9215 with Screw Terminal and NI 9215 with Spring Terminal Safety Voltages Connect only voltages that are within the following limits. Channel-to-COM ±30 V maximum Isolation Channel-to-channel...
Page 4
Do not connect the NI 9215 to signals or use Caution for measurements within Measurement Categories III or IV. NI 9215 with BNC Safety Voltages Connect only voltages that are within the following limits. AI+-to-AI- ±30 V maximum Isolation Channel-to-channel...
Measurement Categories II, III, or IV. Safety Guidelines for Hazardous Voltages You can connect hazardous voltages only to the NI 9215 with screw terminal and the NI 9215 with spring terminal. Do not connect hazardous voltages to the NI 9215 with BNC.
Safety Guidelines for Hazardous Locations The NI 9215 is suitable for use in Class I, Division 2, Groups A, B, C, D, T4 hazardous locations; Class I, Zone 2, AEx nA IIC T4 and Ex nA IIC T4 hazardous locations;...
Page 8
Zone 2 hazardous locations, in ambient temperatures of -40 °C ≤ Ta ≤ 70 °C. If you are using the NI 9215 in Gas Group IIC hazardous locations, you must use the device in an NI chassis that has been evaluated as Ex nC IIC T4, Ex IIC T4, Ex nA IIC T4, or Ex nL IIC T4 equipment.
EMC performance is attained. Physical Characteristics If you need to clean the module, wipe it with a dry towel. For two-dimensional drawings and three- dimensional models of the C Series module and connectors, visit ni.com/dimensions and search by module number.
Page 14
0.2 N · m (1.80 lb · in.) flanges Weight NI 9215 with screw 150 g (5.3 oz) terminal NI 9215 with spring 138 g (4.9 oz) terminal NI 9215 with BNC 173 g (6.1 oz) 14 | ni.com | NI 9215 Getting Started Guide...
Preparing the Environment Ensure that the environment in which you are using the NI 9215 meets the following specifications. Operating temperature -40 °C to 70 °C (IEC 60068-2-1, IEC 60068-2-2) Operating humidity 10% RH to 90% RH, (IEC 60068-2-78) noncondensing...
Grounded Differential Connections Voltage Source – AI– NI 9215 with Screw/Spring Terminal – AI– NI 9215 with BNC 18 | ni.com | NI 9215 Getting Started Guide...
• The NI 9215 with screw terminal and the NI 9215 with spring terminal does not read data accurately if the signal source is outside of the common-mode voltage range.
Page 20
Voltage Source – AI– 100 kΩ NI 9215 with BNC The NI 9215 with BNC has internal circuitry that maintains the common mode voltage range. 20 | ni.com | NI 9215 Getting Started Guide...
Source – AI– NI 9215 with BNC The NI 9215 with BNC has internal circuitry that maintains the common mode voltage range. Connection Guidelines • Make sure that devices you connect to the NI 9215 are compatible with the module specifications.
If your application is subject to high vibration, NI recommends that you follow these guidelines to protect connections to the NI 9215 with screw terminal or the NI 9215 with spring terminal: • Use ferrules to terminate wires to the detachable connector.
NI 9215 Datasheet NI 9215 Datasheet NI-RIO Help NI-DAQmx Help LabVIEW FPGA Help LabVIEW Help RELATED INFORMATION C Series Documentation Services & Resources ni.com/services ni.com/info cseriesdoc Located at ni.com/manuals Installs with the software 24 | ni.com | NI 9215 Getting Started Guide...