Page 4
NI-9250 Getting Started Input signals on each channel are buffered, conditioned, and then sampled ■ by an ADC. Each AI channel provides an independent signal path to the ADC, enabling ■ you to sample all channels simultaneously. AI channels are referenced to earth ground through a protected 50 Ω...
Page 5
The NI-9250 provides an IEPE excitation current for each channel to measure the IEPE sensors. Typical IEPE sensors have a case that is electrically isolated from the IEPE electronics. As a result, connecting the sensor to the NI-9250 results in a floating connection even though the case of the sensor is grounded.
Therefore, the amount of gain or attenuation at a given frequency depends on the data rate. Figure 1. Typical Passband Flatness in DC Coupling for the NI-9250 at the Maximum Data Rate 0.02 0.01...
) controls the data rate (f ) of the NI-9250 . The NI-9250 includes an internal master timebase with a frequency of 13.1072 MHz. Using the internal master timebase of 13.1072 MHz results in data rates of 102.4 kS/s, 51.2 kS/s, 25.6 kS/s, 17.067 kS/s, and so on down to 267 S/s, depending on the decimation rate and the value of the clock divider.
Page 9
When using an external timebase with a frequency other than 13.1072 MHz, the NI-9250 has a different set of data rates. Refer to the software help for information about configuring the master timebase source for the NI-9250 .