3.2.4

Vacuum Monitor (PN 59230) and Pressure Limit Monitor (PN 59706) Gauges - Removal and Installation

Tools required:
7/16" wrench
9/16" wrench
A/R 1/4" wide Teflon tape (PN 60575)
Removal:
1. Remove cover from its enclosure as described in Section 3.2.2 on page 4, and place on a bench top.
2. Disconnect the tubing section that connects to the gauge fitting.
3. Loosen the two thumb-nuts that hold the gauge mounting U-clamp.
4. While gripping the square portion of the gauge port with the 9/16" wrench, remove the female tube
connector (PN 57695) from the gauge.
5. Remove the two thumb-nuts, the mounting U-clamp, and the gauge.
Installation:
1. Before installing a new gauge, wrap two layers of new Teflon tape on the port.
2. Install gauge into panel, secure with U-clamp and tighten the two thumb screws.
3. While gripping the square portion of the gauge port with the 9/16" wrench, tighten the female tube
connector onto the gauge.
4. Attach the tubing section that connects to the gauge fitting.
5. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4.
3.2.5

Pressure and Vacuum Port Quick-Connect Fitting (PN 59997) - Removal and Installation

NOTE: If there is leakage out of the port, replace the test port quick-connect fitting.
Tools required:
1/2" open end wrench
5/8" open end wrench
11/16" open end wrench
A/R 1/4" wide Teflon tape (PN 60575)
A/R 1/2" wide Teflon tape (PN 60911)
Removal:
1. Remove cover from its enclosure as described in Section 3.2.2 on page 4, and place on a bench top.
2. From inside of front panel, disconnect the tubing section that connects to the female tube connector (PN
57695) fitting.
3. Grasp the hex nut at the panel face with a 5/8" wrench and using a 1/2" wrench on the female tube
connector fitting on inside of front panel, turn the female tube connector fitting counter-clockwise and
remove.
4. Grasp the hex nut at the panel face with a 5/8" wrench and using a 11/16" open end wrench on the lock
nut on rear side of panel, turn the lock nut counter-clockwise. Slide quick-connect fitting out of panel
from front.
Installation:
1. Before installing a new Quick-Connect Fitting, wrap two layers of new Teflon tape on the port.
2. Install the quick-connect fitting through front of panel. Thread and tighten lock nut on quick-connect
fitting, from inside of front panel.
3. Thread and tighten female tube connector (PN 57695) fitting into port.
4. Attach the tubing section that connects to the female tube connector fitting.
5. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4.
6
UVC1000/UVC1010 Operation and Maintenance Manual