Condec UVC1000 Operation And Maintenance Manual

Table top & rack-mountable pneumatic vacuum generator / pressure controller

Advertisement

Quick Links

UVC1000/UVC1010
Table Top & Rack-mountable Pneumatic
Vacuum Generator/Pressure Controller
Operation and Maintenance
Manual
Condec Sales Phone Number: (888) 295-8475
Condec Web Site: www.4condec.com
69481

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the UVC1000 and is the answer not in the manual?

Questions and answers

Summary of Contents for Condec UVC1000

  • Page 1 UVC1000/UVC1010 Table Top & Rack-mountable Pneumatic Vacuum Generator/Pressure Controller Operation and Maintenance Manual Condec Sales Phone Number: (888) 295-8475 Condec Web Site: www.4condec.com 69481...
  • Page 2: Table Of Contents

    Test Procedure ............4 3.2.2 UVC1000/UVC1010 Cover - Removal and Installation ......4 3.2.3 Pressure Limit Control (PN 58409) - Regulator Removal and Installation .
  • Page 3: About This Manual

    Introduction The UVC1000/UVC1010 offers a combination of features, performance, versatility and reliability. Some of the more outstanding features are listed below: • Two models available: table top and 19" rackmount. • Portable: These compact, self-contained systems are easily carried and operated by only one person. Total weight is approximately 7 pounds.
  • Page 4 The following schematic provides an overview of the UVC1000/UVC1010’s function. Figure 1-1. UVC1000/UVC1010 Flow Diagram UVC1000/UVC1010 Operation and Maintenance Manual...
  • Page 5: Operation

    & input port. Figure 2-1. Input Operation Procedure NOTE: UVC1000 shown, UVC1010 Rack Mount unit also has an input port on back of unit. Pressure Measurement Sequence To prepare for actual calibration usage, see Figure 2-1 above and proceed as follows: 1.
  • Page 6: Maintenance

    This section outlines the test and mechanical repair procedures for the UVC1000/UVC1010 Vacuum Generator/Pressure Controller. The repair procedures cover the major components and subassemblies which are critical to the proper functioning of the UVC1000/UVC1010 and that will likely need periodic maintenance over the life of the unit.
  • Page 7: Pressure Limit Control (Pn 58409) - Regulator Removal And Installation

    2. Insert the new regulator into the panel through rear of panel and rotate to proper position. Thread and tighten the mounting nut onto the regulator body from the panel front. 3. Attach the tubing sections to the inlet and outlet fittings. 4. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. Maintenance...
  • Page 8: Vacuum Monitor (Pn 59230) And Pressure Limit Monitor (Pn 59706) Gauges - Removal And Installation

    3. While gripping the square portion of the gauge port with the 9/16" wrench, tighten the female tube connector onto the gauge. 4. Attach the tubing section that connects to the gauge fitting. 5. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. 3.2.5 Pressure and Vacuum Port Quick-Connect Fitting (PN 59997) - Removal and Installation NOTE: If there is leakage out of the port, replace the test port quick-connect fitting.
  • Page 9: Input Port (Pn 59761) - Removal And Installation

    9. Tighten the AN bulkhead end nut approximately 1-1 /4 turns beyond finger tight, using 9/16" wrench. 10. Attach the tubing section that connects to the male tube connector (PN 57705) fitting. 11. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. 3.2.7 Input Port Filter (PN 54188) - Removal, Cleaning and Installation The port filter is a sintered element filter that can be easily removed for inspection and cleaning.
  • Page 10: Vacuum Generator (Pn 57960) - Removal, Cleaning And Installation

    (PN 59761), panel. 4. Tighten the AN bulkhead end nut approximately 1-1 /4 turns beyond finger tight, using 9/16" wrench. 5. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. 3.2.8 Vacuum Generator (PN 57960) - Removal, Cleaning and Installation The vacuum generator is a Venturi device which creates a vacuum flow by use of a regulated air flow.
  • Page 11: Specifications

    Specifications Models: Pressure Hose Fittings: UVC1000 (PN 56085): Table Top Unit Quantity Supplied: Two; one each for Pressure and UVC1010 (PN 59415): 19” Rack Mount Unit Vacuum port fittings. Style: Quick-disconnect male one end, 1/8 NPT female on the other Operating Temperature:+40°...
  • Page 12: Uvc1000/Uvc1010 Warranty And Return Policy

    UVC1000/UVC1010 Warranty and Return Policy If possible, please save original packing material which is specifically designed for the unit. Should it be necessary to ship the unit back to the factory, a suitable shipping container must be used along with sufficient packing material.
  • Page 13: Uvc1000/Uvc1010 Return Material Authorization Form

    ) NO SHIP TO Address: COMPANY NAME: STREET: CITY, STATE, ZIP: ATTN: CONDEC • 3 SIMM LANE • DOOR D, UNIT 2A • NEWTOWN, CT 06470 ATTN: PRESSURE PRODUCTS/REPAIR LAB TEL: 888-295-8475 • FAX: 203-364-1556 or 715-234-6967 WEB SITE: www.4condec.com...

This manual is also suitable for:

Uvc1010

Table of Contents