Introduction The UVC1000/UVC1010 offers a combination of features, performance, versatility and reliability. Some of the more outstanding features are listed below: • Two models available: table top and 19" rackmount. • Portable: These compact, self-contained systems are easily carried and operated by only one person. Total weight is approximately 7 pounds.
Page 4
The following schematic provides an overview of the UVC1000/UVC1010’s function. Figure 1-1. UVC1000/UVC1010 Flow Diagram UVC1000/UVC1010 Operation and Maintenance Manual...
& input port. Figure 2-1. Input Operation Procedure NOTE: UVC1000 shown, UVC1010 Rack Mount unit also has an input port on back of unit. Pressure Measurement Sequence To prepare for actual calibration usage, see Figure 2-1 above and proceed as follows: 1.
This section outlines the test and mechanical repair procedures for the UVC1000/UVC1010 Vacuum Generator/Pressure Controller. The repair procedures cover the major components and subassemblies which are critical to the proper functioning of the UVC1000/UVC1010 and that will likely need periodic maintenance over the life of the unit.
2. Insert the new regulator into the panel through rear of panel and rotate to proper position. Thread and tighten the mounting nut onto the regulator body from the panel front. 3. Attach the tubing sections to the inlet and outlet fittings. 4. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. Maintenance...
3. While gripping the square portion of the gauge port with the 9/16" wrench, tighten the female tube connector onto the gauge. 4. Attach the tubing section that connects to the gauge fitting. 5. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. 3.2.5 Pressure and Vacuum Port Quick-Connect Fitting (PN 59997) - Removal and Installation NOTE: If there is leakage out of the port, replace the test port quick-connect fitting.
9. Tighten the AN bulkhead end nut approximately 1-1 /4 turns beyond finger tight, using 9/16" wrench. 10. Attach the tubing section that connects to the male tube connector (PN 57705) fitting. 11. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. 3.2.7 Input Port Filter (PN 54188) - Removal, Cleaning and Installation The port filter is a sintered element filter that can be easily removed for inspection and cleaning.
(PN 59761), panel. 4. Tighten the AN bulkhead end nut approximately 1-1 /4 turns beyond finger tight, using 9/16" wrench. 5. Test UVC1000/UVC1010 as described in Section 3.2.1 on page 4. 3.2.8 Vacuum Generator (PN 57960) - Removal, Cleaning and Installation The vacuum generator is a Venturi device which creates a vacuum flow by use of a regulated air flow.
Specifications Models: Pressure Hose Fittings: UVC1000 (PN 56085): Table Top Unit Quantity Supplied: Two; one each for Pressure and UVC1010 (PN 59415): 19” Rack Mount Unit Vacuum port fittings. Style: Quick-disconnect male one end, 1/8 NPT female on the other Operating Temperature:+40°...
UVC1000/UVC1010 Warranty and Return Policy If possible, please save original packing material which is specifically designed for the unit. Should it be necessary to ship the unit back to the factory, a suitable shipping container must be used along with sufficient packing material.