All the products and solutions for cleaning and disinfection must be used according to the manufacturer’s instructions. The recommendations for cleaning, disinfection and sterilization are updated and available on MORIA’s website: http://www.moria-surgical.com MORIA recommends to: • avoid touching potentially contaminated areas,...
Page 2
Discard the disposable head and aspiration tubing after each procedure in an appropriate container. Warning: The reuse of single-use products, or the use of consumables other than those supplied by MORIA, may entail serious surgical consequences for the patient and damage the microkeratome.
Page 3
III) CLEANING-DISINFECTION-STERILIZATION-STORAGE OF THE REUSABLE SUCTION RINGS This protocol has been approved by the French XV/XX National Ophthalmological Hospital (CHNO) (Paris, France), Hygiene/ Sterilisation Department (Dr. Patrice RAT). Step 5: Double cleaning On leaving the operating theatre, immediately immerse the devices in an aldehyde-free detergent/disinfectant solution for at least 15 minutes according to manufacturer recommendations.
Page 4
As only MORIA and its agents are fully expert in MORIA products, servicing and maintenance must be carried out by MORIA or its approved agents. MORIA shall not be held liable for any malfunction or damage to the apparatus, poor results, or surgical complications due to maintenance having been carried out by an unqualified operator or third party.