Table of Contents Chapter1 Brief Introduction ......................5 Chapter 2 Functions ........................6 Chapter 3 Main Technical Parameters of KS833 Series...............7 Main Technical Parameters of KS833 (Grade 0.05)..........7 3.2 Main Technical Parameters of KS833 (Grade 0.1)............9 3.3 KS833 Operation Panel ....................11 3.4 System Configuration....................12...
Page 3
KS833 Instruction Manual Attachment I ..........................73 Attachment Ⅱ ...........................74 KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 4
3. When measuring 2-element active, reactive and power, Ub and Uo should be shorted. 4. No non-technical person should be allowed to enter the place when conducting calibration using the standard source, or the instrument output precision will be affected. KINGSINE ELECTRIC AUTOMATION CO., LTD.
PAM (power amplifier module), large LCD displayer, and imported components and ICs. With its advanced technology, multi-function, small and portable size, with its superb performance and strong function, KS833 has been welcomed by our customers. KINGSINE ELECTRIC AUTOMATION CO., LTD.
2. soft touch keyboard operation, and 3. operation under Windows on PC; l) Storage capacity up to 1,000 calibrated meters’ data. m) Through software you can round off the calibrated data, print verification report and calibration data. KINGSINE ELECTRIC AUTOMATION CO., LTD.
KS833 Instruction Manual Chapter 3 Main Technical Parameters of KS833 Series 3.1 Main Technical Parameters of KS833 (Grade 0.05) a) Output & measure AC voltage Available steps: 10V, 30V, 100V, 300V, 750V; automatic switch is possible between the steps. Adjustment range: Step x (0 – 120)% Adjustment fineness: Step x 0.01%...
Page 8
Max output load capacity of DC voltage & current Voltage output 20VA, current output 25VA n) Measure DC: Voltage measuring: ±10V Current measuring: ±20mA Accuracy: 0.01%RG o) Reference conditions for parameter test: Ambient temperature: 22º± 1º Working temperature: 0º~ 40º, humidity: ≤85% KINGSINE ELECTRIC AUTOMATION CO., LTD.
KS833 Instruction Manual Working voltage range: 220VAC±15%,50Hz 3.2 Main Technical Parameters of KS833 (Grade 0.1) a) Output & measure AC voltage Available steps: 10V, 30V, 100V, 300V, 750V; automatic switch is possible between the steps. Adjustment range: Step x (0 – 120)% Adjustment fineness: Step x 0.01%...
Page 10
Voltage output 20VA, current output 25VA n) Measure DC: Voltage measuring: ±10V Current measuring: ±20mA Accuracy: 0.02%RG o) Reference conditions for parameter test: Ambient temperature: 22º± 1º Working temperature: 0º~ 40º, humidity: ≤85% Working voltage range: 220VAC±15%,50Hz p) Size: KINGSINE ELECTRIC AUTOMATION CO., LTD.
KS833 Instruction Manual 3.4 System Configuration KS833 Comprehensive Calibrating Unit for Electrical Measuring 1 unit Instruments - host Operating software for KS833 Comprehensive Calibrating Unit for 1 set Electrical Measuring Instruments Cable 1 cable Test wire for power industry 1 parcel...
4.2.1 Brief Introduction Six items are displayed on the main menu, you can select them by the following steps: press direction keys or turn the Rotary Encoder to move the cursor on the button you want, KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 14
“Instrument Calibration” button to enter KS833’s function module of instrument calibration; press down “Standard Source Calibration” button to enter KS833 standard source calibration; press down “Standard Source” button to enter KS833 standard source; press down “Connected test” to connect to host (upper) computer, which then controls the unit;...
Chapter 5 Multifunction Standard Source 5.1 Brief Introduction KS833 multifunction standard source is composed of six parts: DC Standard Source, Single Phase Standard Source, Phase Standard Source, Three-phase Standard Source, Power Standard Source and Harmonic Standard Source; the interface is shown as below in...
Page 16
You can click OK in the dialogue box to stop the alarm. You can restart the test after eliminating the fault. 5.2.2.9 Stop Test When output is not needed from the instrument, you can click Stop Test button to stop, KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 17
Single Phase AC Standard Source, see Figure 6 for its interface (Phase A output) 5.2.3 Figure 6 KS833 Single Phase AC Standard Source interface 5.2.3.1 Same test procedure as that of DC Standard Source. 5.2.3.2 The text box of Output Port will display the amplitude of output voltage or current.
Page 18
KS833 Instruction Manual Figure 7 KS833 AC Phase Standard Source interface 5.2.4.1 Select Operation Mode Click the “Operation Mode” dropdown box, select Ua Ub Phase, Uala Phase or lalb Phase positions. The instrument will output the voltage or ampere at the corresponding output terminal, the difference between two phases will be the phase value you need.
Page 19
Of which the Ua output port cannot be changed, and it is always effective. Figure 8 KS833 Three-phase AC Standard Source interface 5.2.5.2 Change Working Step Click Enter on the dropdown box of Working Step to select the working step you want, the selected Step is effective only for the corresponding output port, and different output ports and output simultaneously under different working steps.
Page 20
5.2.5.9 On-line Adjustment: identical as 5.2.4.10 5.2.5.10 Fine Adjustment Move cursor to the edit boxes of amplitude or phases, then turn the selection encoder, the figures will increase or decrease by step length of its minimum position. 5.2.5.11 Overload Warning KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 21
Real Reactive Power”, “3 Element 90° Reactive Power”, “2 Element 90° Reactive Power.” Figure 10 KS833 Power Standard Source interface 5.2.6.2 Change Working Step Click enter on the dropdown box of Volt step and Amp step, select the working step you want (the standard source provides 3 volt steps: 750V, 300V, 100V;...
Page 22
5.2.6.10 Set Stepping KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 23
& reactive power of all elements. Figure 13 KS833 Output Power Display Interface 5.2.6.12 Auto Switch & Manual Switch Click Auto Switch button to toggle between manual switch and auto switch. When you enter Manual Switch, you need to switch amp steps to enhance the precision of power output.
Page 24
Harmonic Overlapping Standard Source: see Figure 14 for its interface Figure 14 KS833 Harmonic Overlapping Standard Source interface 5.2.7.1 Select Fundamental Step Click Fundamental dropdown box to select from the steps, the standard source provides five volt steps: 750V, 300V, 100V, 30V and 10V; and five amp steps 25A, 10A, 5A, 1A and 100mA.
Page 25
Click Stop Test during test, the instrument will close output. At the same time, output display will be switched to Set Harmonic, the harmonic component table will become harmonic overlapping input component table and display the set value of each harmonic component of the selected harmonic output port. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Code list for the desired Chinese character and input it as the code specified, both temperature and humidity are ranged between 0 and 100. Figure 16 KS833 Set Working Environment Interface KINGSINE ELECTRIC AUTOMATION CO., LTD.
KS833 Instruction Manual Figure 17 KS833 Set Working Environment 6.2 Print Report Enter Print Report interface, first select instrument type, then input Model and Number, if previously stored data of this instrument is available the system will automatically call it out, finally, click Print button;...
Figure 20 KS833 Instrument Attributes – Attention interface 6.3 Input Instrument Attributes Figure 21 shows the input interface of KS833 Instrument Attributes, the user may select instrument name and type and input other information displayed, then press down “Range Input” button to enter corresponding instrument input interface; after entering all data then press down “OK”...
KS833 Instruction Manual Figure 21 KS833 Instrument Attributes interface Figure 22 is the setting of Mutual Inductance Parameters PT 1 indicates primary voltage, PT 2 indicates secondary voltage; CT 1 indicates primary current and CT 2 indicates secondary current. “PT 1 is 66000”...
Page 30
Figure 24 shows Power Range Input interface, the inputting method is similar with that of AC/DC Range Input. Figure 24 Power Range Input interface 6.4.3 Phase Range Input Figure 25 shows Phase Range Input interface, the inputting method is similar with that of AC/DC Range Input. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Down tests. After that, the cursor will move up one line after each point calibration is completed. During the test, calibrating one point will: 1). First display the value of the calibrated meter; 2). Show the Up value or Down value of the standard meter; 3) KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 32
The operation is similar with complete check range calibration. During the calibration, the work after reading each number is the same with complete check calibration except that the variation will not be calculated. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 33
Figure 28 & 29. Figure 28 is for Complete Check - Range Calibration, Figure 29 is for Incomplete Check – Range Calibration. All the operations are the same with those of Portable AC voltmeter and ammeter. Figure 28 DC Voltage / Current Complete Check Range Calibration KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 34
“Next” to enter the incomplete check range calibration interface shown in Figure 31; press “Print” to print all calibration data and result; press “Show Status” to display output status; press “Pg Up” to move up one page; press “PgDn” to move down one page. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 35
“OK” to save calibration data and return to complete check range. Figure 31 Phase Incomplete Check – Range Calibration 6.5.1.4 Portable Frequency Meter (Phase A Output) Figure 32 Frequency Complete Check – Range Calibration KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 36
6.5.1.5 Calibrate Power Meter 1. On the KS833 Instrument Attributes page, click OK to enter the page of KS833 power meter complete check range calibration. Figure 34 KS833 Power Meter Complete Check Range interface 2. Complete Check – Range Calibration A) According to the instrument range parameters input in Instrument Attributes page, the system will first enter complete check range (also called basic range).
Page 37
Adjustment Fineness at any time, click Enter to select different adjustment fineness values. Figure 35 KS833 Power Meter Complete Check Range B) Select the power factor you want to calibrate, click Start Test button. Figure 36 KS833 Power Meter Complete Check Range When the power factor is 1.0, both up value and down value need to be...
Page 38
Figure 37 KS833 Power Meter Complete Check Range conclusion interface After the instrument completes testing one verification point, the user may click Conclusion button to view the max. error, max. variation and max.
Page 39
3. Incomplete Check – Range Calibration A) According to the power incomplete check ranges input by the user, the system will calibrate each range one by one. Figure 39 KS833 Power Meter Incomplete Check Range KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 40
6.5.1.6 Calibrate Low Power Factor Meter On the KS833 Instrument Attributes page, click OK to enter the page of KS833 power factor meter complete check range calibration. Figure 41 KS833 Power Factor Meter Complete Check Range Complete Check –...
Page 41
Down cells; if it’s a digital meter, then the range calibration is completed here. iii. The rest process is the same with that of power meter calibration. Incomplete Check – Range Calibration Figure 42 KS833 Power Factor Meter Incomplete Check Range KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 42
6.5.3.1 For experimental digital AC voltmeter and ammeter, Figure 43 shows the interface of complete check range calibration, Figure 44 shows incomplete check range calibration. Figure 43 Experimental Digital AC Complete Check Range Calibration interface KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 43
The calibration process of experimental digital DC voltmeter and ammeter are basically the same with that of portable DC voltmeter and ammeter, only that for digital meters there’s no need to calibrate variation. Figure 45 Experimental Digital DC Voltmeter and Ammeter Complete Check Range Calibration interface KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 44
Figure 47 & 48. Figure 47 is the interface of complete check range calibration; Figure 48 is the interface of incomplete check range calibration. Figure 47 Experimental Digital Phase Complete Check Range Calibration interface KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 45
Calibration. The calibration process of digital frequency meter is basically the same with that of portable frequency meter. The only difference is that for digital frequency meters there’s no need the check variation. Figure 49 Experimental Digital Frequency Complete Check Range Calibration interface KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 46
Experimental Digital Frequency Incomplete Check Range Calibration interface 6.5.4 Field Digital Meter Field digital meters are also divided into two types: with or without mutual inductance, the calibration process is corresponding with that of experimental digital meters. KINGSINE ELECTRIC AUTOMATION CO., LTD.
52, select Energy Meter in Select Instrument, then input other parameters, click OK to enter the input interface of energy meter. (Note: Model and Number are required to fill out) Figure 52 KS833 Instrument Attributes 7.1.1 Input Parameters For Energy Meter...
Page 48
Inductive Portable Meter, Inductive Installation Meter, Electronic Meter - Type A, Electronic Meter – Type B, Electronic Installation Meter. You can select the operation mode and type you want in these dropdown menus. Figure 54 Combo Box of Operation Mode KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 49
7.1.3 Set the parameters required in Figure 53, click OK to enter the dialogue box of energy meter calibration, as is shown in Figure 57, you can select and enter any of the items. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Figure 55, if you change the working current from 0.50% IB to 1% lb, the test duration will change to 545 sec. Click Start Test and confirm that, after the test it will prompt whether the booting test of energy meter is passed or not. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Select Basic Error in Figure 57 and press Enter to enter the interface of Balanced Load Test of Energy Meter, see Figure 60. The power factor should be set in Figure 53 (energy meter - power factor to be calibrated). Figure 60 Balanced load test of energy meters KINGSINE ELECTRIC AUTOMATION CO., LTD.
Test, press Enter to confirm, the system will prompt that it’s closing the loop, at the same time the test duration is shortening, till it reaches the set duration the test will be over, and test result will be displayed. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Balanced Load Test will display “Save”, you click “Save” and it will automatically changed into “Print”, plug the printer, test result will be printed. (Note: For auto tests, the result cannot be saved or printed.) KINGSINE ELECTRIC AUTOMATION CO., LTD.
8.1 Operation Flow Figure 63 Transducer Calibration interface 8.1.1 In the main interface of KS833 Electrical Measuring Transducer Calibration, click “Calibrate Transducer Output” button to enter the calibration interface of calibrating transducer output nominal value. Click “Calibrate Transducer Measuring” button to enter instrument calibration page.
8.2.3 Select Working Step 8.2.3.1 Click “Working Step” combo box; select the working step to be calibrated. Three steps for DC Voltage: 1V, 5V, 10V; and three steps for DC Current: 1mA, 10mA, 20mA. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Save Data button to save the calibration data. Before exit this page, the system will automatically save all the unsaved calibration data of the cession. (Note: there will not be prompt to save, the saving process is mandatory.) 8.4 Verify Transducer Measuring 8.4.1 Instrument Parameter Setting KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 57
8.4.2.4 If you want to select other instrument to be calibrated, click “Back” to return to the Instrument Parameter Setting page. Click “Next” to enter Instrument Inspection page, to do inspection test. Click “Cancel” to cancel this cession of transducer inspection and return to Main Menu. 8.4.3 Transducer Inspection Test KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 58
Figure 70 Example 1 Example 2: Frequency Transducer Verification, set the output signal as bi-directional voltage, 7 verification points, for this meter to be verified they are: -100, -66.67, -33.33, 0.00, 33.33, 66.67, and 100. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 59
12 verification points means 12 points for each power factor. (Note: Pre-processing Duration is only applicable when the verification mode is Auto.) KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 60
When the output signal is bi-directional, under the column of Calibrated Meter there appears +0 or -0, which indicates that the current is zero at the moment, and the powers factors are ±1, ±0.5L, ±0.5C respectively. KINGSINE ELECTRIC AUTOMATION CO., LTD.
8.7 Data Processing After Stop Test, you can click “Conclusion” to view whether the tested instrument passed or not, click “Print” to print the file, click “Save” to save the data of this test cession. KINGSINE ELECTRIC AUTOMATION CO., LTD.
Chapter 9. Calibrate Multifunction Standard Source 9.1 Brief Introduction KS833 multifunction standard source calibration is divided into five parts: AC standard source calibration, phase standard source calibration, DC standard source calibration, power standard source calibration and change password; the interface is shown in Figure...
Page 63
Working Step dropdown box. Figure 78 KS833 AC Standard Source Calibration interface 9.2.2.2 Connect the standard source meter to the selected the output port. 9.2.2.3 Click Start Test button, wait till system output is stable, and then input the indicating value of the standard source in the edit box of the popup dialogue box.
Page 64
9.2.3.2 When calibrating phase standard source, select the phase output terminal in the dropdown box of Operation Mode, Figure 79 KS833 AC Phase Standard Source Calibration interface Select all voltage steps in Working Step dropdown box. 9.2.3.3 Connect the phase meter to the corresponding output port, if the test phases are those or Ua and Ub or Uc, then the two phases’...
Page 65
KS833 Instruction Manual Figure 80 KS833 DC Standard Source Calibration interface 9.2.4.2 Click the “Operation Mode” dropdown box, select “DC Voltage” or “DC Current.” In the Working Step dropdown box, select the working step you want to calibrate, if voltage is selected for the operation mode, voltage steps will be display, and current steps will be displayed if current is selected.
Page 66
KS833 Instruction Manual Figure 81 KS833 AC Phase Power Standard Source Calibration interface 9.2.5.2 Select the calibration options in the dropdown box of Operation Mode: 3P4W, 3P3W or Cross Phase Input; select the calibration voltage step in the Voltage Step dropdown box and select the current step in the dropdown box of Current Step;...
Page 67
Old Password; second, you input different New Passwords. So you should be careful when you change password 9.3 Attention: Calibration data is the data necessary for the instrument to work normally, so in calibrating standard source, you should make sure the correctness of each calibration point. KINGSINE ELECTRIC AUTOMATION CO., LTD.
10.1 Functions: 10.1.1 Connected test is used to communicate with the test software in host computer and exercise control of KS833 standard source. 10.1.2 Under the control of host computer software, directly calibrate the instrument’s output, calibration and connected closed loop calibration, as well as the sampling of actual output.
Page 69
KS833 Instruction Manual Figure 84 KS833 Connection spirit interface Figure 85 KS833 Connection spirit interface 10.2.4 When the connection is established, the system will be directly controlled by the host computer software, without any intervention from the user. 10.2.5 The user may click Disconnect at any time to disconnect with the host computer. If you want to disconnect during a test process, the system will automatically close test output and then disconnect with the host computer.
1. Voltage content of the h harmonic × In which: —— voltage of the h harmonic (root-mean-square) ——Voltage of the fundamental (root-mean-square) 2. Current content of the h harmonic × In which: —— Current of the h harmonic (root-mean-square) KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 71
KS833 Instruction Manual ——Current of the fundamental (root-mean-square) 3. Voltage content of the harmonic ∞ ∑ 4. Current content of the harmonic ∞ ∑ 5. Voltage total harmonic distortion × 6. Current total harmonic distortion × % KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 72
KS833 Instruction Manual KINGSINE ELECTRIC AUTOMATION CO., LTD.
Page 73
3.Software upgrading: After connection established successfully, click Send File button or menu, wait till a dialogue box pops up, select the latest program for KS833 Comprehensive Calibrating Unit for Electrical Measuring Instruments provided by Kingsine Electric Automation Co., Ltd.
Page 75
KS833 Instruction Manual StdVal standard value total harmonic distortion Trans transducer variation verify Volt voltage KINGSINE ELECTRIC AUTOMATION CO., LTD.