Summary of Contents for Vapotherm Heliox Precision Flow
Page 1
This guide provides you with basic instructions on how to set up and operate the Precision Flow® Heliox. Before operating the Vapotherm Precision Flow® Heliox, please review the Instructions For Use which can be found at our website www.vapotherm.com and on the USB drive provided in the Start-Up Kit.
Page 2
ALARM ICON WARNING INDICATES CAUSE ACTION Check gas supply. If not corrected, disconnect patient. GENERAL FAULT Malfunction of Internal Unplug AC power, press and sensor or control – – IN FLOW component failure hold Run/Standby button for system (FLASHING) 3 seconds to clear the alarm, send for service.
Page 3
ALARM ICON WARNING INDICATES CAUSE ACTION Disconnect patient. Remove RUN mode: faulty disposable patient circuit. Cartridge and/or DPC not sensor or cartridge Check cartridge installation. detected. Unit will not run. not detected. Check sensor windows are clean. Disconnect patient. Place CARTRIDGE Gas bubbles in water Excessive gas...
Page 4
VAPOTHERM, INC. 0297 100 Domain Drive T: 855 557 8276 Domestic RMS-UK Ltd. Exeter, NH 03833 +1 (603) 658-5121 International 28 Trinity Road T: 603-658-0011 ts@vtherm.com Nailsea, North Somerset BS48 4 NU United Kingdom May be patented. www.vapotherm.com/patents 3101541 Rev. A...