Connecting electrical equipment to the Multi-Socket Outlet effectively leads to creating a Medical Electri-
cal system and can result in a reduced level of safety
An extension cord or additional Multi-Socket Outlet may not be connected to the T7's integrated
Multi-Socket Outlet
Only equipment to be used as part of the T7 system or compatible with power requirements may
be connected to the Multi-Socket Outlet for power
Potential loss of data if battery runs out of power. Plug in AC cable to charge battery if the low-battery
alarm sounds
The Non-powered T7 cart may have greater risk of leakage current with user-installed technology
on-board
Installing technology in the Non-Powered T7 will pose a greater risk of hazard due to liquid spills.
Extra precaution should be taken.
A hazardous risk of tipping exists when the T7 encounters either of the following heightened concern
conditions:
• Navigating a step up or down of 40mm (1.5") or greater; proceed with caution or find assistance in moving the
cart over a change of height of this magnitude.
• Positioned on an inclined surface of 10° or more; avoid when applicable, but if necessary, proceed with
caution or find assistance to maintain cart stability by firmly holding push handle at all times.