Service Manual - PHP 1000 Pump
T2 Pressure calibration (For PHP 1000 pressure calibration
procedure)
SOP:
Comfort 9 & Comfort 10 LED
STEP-9 : Press the Alarm mute button, the LED
lights of Comfort 1 and Comfort 2 will
turn off. After that, the LED lights of
Comfort 9 and Comfort 10 will light up
and flash, and the compressor will stop
working. It means the low pressure
calibration is finished, and then proceed
to perform a high pressure calibration.
60mmHg
STEP-10:Provide a stable 60 mmHg pressure.
2008.04 V1.0
19