Service Manual - PHP 1000 Pump
T2 Pressure calibration (For PHP 1000 pressure calibration
procedure)
SOP:
Comfort 1 & Comfort 2 LED
STEP-7:Press the alarm mute button, the low
pressure LED will turn off. The LED lights of
Comfort 1 and Comfort 2 will light up and
flash, and the compressor will stop
working. It means the pulling air tubing
calibration was finished, and then it could
begin to perform a low pressure
calibration.
30mmHg
STEP-8:Provide a stable 30 mmHg pressure.
2008.04 V1.0