Bowles Corporation
Operation
When operated according to the instructions given in this chapter, the UW-7 digital
Ultrasound Wattmeter will quickly and accurately measure the energy output of
ultrasound devices.
Operating Environment
The UW-7 must be allowed to adjust to room temperature before the unit can be
operated. The relative humidity in the room should not exceed 90%.
Refer to Chapter 4 for additional information on safety, storage and shipping
requirements for the UW-7.
Precautions
When using the UW-7 to test ultrasound devices, follow the safety precautions
outlined by the ultrasound manufacturer.
Do not use any coupling medium other than degassed and distilled water.
Use of improper media can lead to reading variations of up to 10%.
Do not spill water on the front or rear panel.
Do not apply more than 30 W of input power to the UW-7.
Do not operate the UW-7 for more than 1 hour during testing. Change water
after 45 minutes of use.
Remove and dry the cone when the test is complete.
Drain the unit completely after each test session.
UW 7
25