Temperature resistance
Cable entries
Measured variable
Measuring range
(application)
Input signal
Measuring frequency
Galvanic isolation
Switch behaviour
Power-on behaviour
Fail-safe mode
Switching delay
Ex specifications
4
Cable specifications
Use a shielded cable in the event of strong electromagnetic radiation.
The connecting cables must withstand an ambient temperature of +20 K.
M20x1.5 (cable gland); NPT ½; G ½
Input
Level (according to the mounting location and the overall length)
The measuring range depends on the mounting location of Soliphant T and the length of the pipe extension
selected. The pipe extension is available in the following lengths: 500 mm, 1000 mm, 1500 mm (20 in, 40 in,
60 in).
Probes covered => small amplitude
Probe not covered => large amplitude
700...800 Hz
Output
FEM22:
Between sensor and power supply
FEM24:
Between sensor, power supply and load
Binary
When switching on the power supply the output is set to "signal on alarm".
After a maximum of 3 s it switches to the correct output signal.
Minimum/maximum quiescent current safety can be switched at electronic insert
MAX = maximum safety:
When the vibrating rod is covered, the output switches in the direction of the signal on alarm
Used for overfill protection for example
MIN = minimum safety:
When the vibrating rod becomes exposed, the output switches in the direction of the signal on alarm
Used for empty running protection for example
0.5 s when the sensor is covered
1 s when the sensor is exposed
FEM22, FEM24:
– Explosion protection for explosive dust-air mixtures:
Dust-Ex, DIP
Soliphant T FTM20, FTM21
Endress+Hauser