10.

Internal calibration

Internal calibration is performed automatically after each start-up, also every 2 hours
and with temperature changes more than 1°C.
To calibrate the balance with internal weight, simply empty the pan and press
CALIBRATION
Internal calibration in progress:
Please wait ...
CALIBRATION
Internal calibration in progress:
Taring, please wait ...
CALIBRATION
Internal calibration in progress:
Loading, please wait ...
Max ...
Min ...
0%
In case internal calibration does not ensure proper accuracy of the balance (e.g. results
of weighing object of known weight are wrong), please contact the nearest service
centre.
USER MANUAL
e= ...
d= ...
0,000
g
100%
Press
key.
Until calibration process is finished do
not perform any operation, as any
vibrations and shocks may affect the
process of calibration and delay the
calibration or deteriorate the result.
The calibration weight is placed on the
pan three times to avoid inaccurate
calibration result. In case any problems
during calibration, the error is signalised
on the display and calibration process is
stopped. The result of correct calibration
is zero indication.