Batteries
Battery operating life
The battery operating life depends on the type of battery, actual usage
and environmental factors.
A new model 9145 battery provides:
A minimum of 12 hours (14 hours typical) of device operating
time at 20°C to 30°C ambient temperature with no shocks
delivered, OR
A minimum of 9 hours of device operating time at 0°C ambient
temperature with no shocks delivered, OR
A minimum of 9 hours of device operating time at 50°C ambient
temperature with no shocks delivered, OR
up to 290 shocks (typical)
Table 2-1: Normal battery life
Model
9145 lithium
sulphur dioxide
9144 lithium ion
Battery shelf life
The batteries have a shelf life of five years from the date of
manufacture. Shelf life is defined as the length of time a battery can be
stored at room temperature, prior to installation into AED, and still
meet the specifications under battery operating life.
Note: Storing the battery outside its specific range (0-50°C) will
decrease battery life.
Powerheart® AED G3 Pro 9300P
Estimated Shelf
Life (from date of
manufacture)
5 years
5 years
Full Operational
Replacement Guarantee
(from date of installation)
1 year from date of
installation or 12 hours of use,
whichever occurs first
2.5 years or 300 battery
charge/discharge cycles,
whichever comes first
70-00969-01 F
Typical
Shocks
up to 290
Per charge:
60 shocks
minimum
(100 shocks
typical)
2-7