Procedure for purging remaining refrigerant from this machine. 
 
1.
Close the ports of the system being serviced that are connected to the 
input port of the machine. 
2.
Turn off the  recovery machine. 
3.
Turn the Input valve to the PURGE position. 
4.
Turn the MODE valve to the PURGE position. 
5.
Restart the machine. 
6.
Run until desired vacuum is achieved. 
7.
Close the ports on the recovery tank and the machine. 
8.
Turn the machine off. 
9.
Return the MODE valve to the RECOVER position. 
10.
Disconnect and store all hoses and dry filter 