iQUBE
Junction Box
Quick-Start Guide
Use this document to install the iQUBE
for an 8-cell truck scale that is being used with two iQUBE
Load cells or simulators must be connected before using the instructions in this document. See the iQube2
installation manual (available on the Rice Lake website at www.ricelake.com).
RS-422 and RS-485 are used interchangeably. TEDS is not supported.

Configuration using 920i

Establish a Connection

1. Connect the iQUBE
RS-422 is recommended due to the 115,200 baud rate and a maximum distance of 1000' at this high baud
rate. RS-232 has a maximum distance of 10'.
Table 1 indicates the connections for RS-422 communications between a
Two-wire half duplex is available on Port 4 of the
2. Set
dip switches on CPU board to 485 (both set to
S2
locations.
If a change is made to switch position, cycle power on the iQUBE
read on power-up.
3. Once power is on again, place
4. Enter Config mode on the
5. Change the
PORTTYPE
SCALES
February 07, 2017
2
using Rice Lake Weighing Systems recommended settings. It is written
2
®
and the
using an RS-422 connection.
920i
920i Board
J10 Connector (Port 4)
RS-422 Signal
GND
RS-422 A
RS-422 B
Table 1. RS-422 – 2-Wire Connections to 920i
in the
SW1
and navigate to the
920i
parameter to
422.
SERIAL
FEATURE
SCALE
IND SC
DISPLAY
Figure 1. PORTTYPE Parameter
2
CPU boards.
.
920i
iQUBE
J7 Connector
Pin
Pin
RS-422 Signal
1
1
5
4
RS-485 A
6
5
RS-485 B
). See Figure 10 on page 6 for
OFF
position.
OFF
under the
PORTTYPE
PFORMT
SETPTS
DIG I/O
PORT 4
IQUBE2
PROGIN
CONFIG
COMM SEL
and the iQUBE
920i
2
GND1
2
while in Setup mode, switch position is
menu.
SERIAL
VERS
CMD
PORTTYPE
UPDATE
232
422
485
2
.
dip switch
S2
PN 119611 Rev A

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the iQUBE2 and is the answer not in the manual?

Questions and answers

Summary of Contents for Rice Lake iQUBE2

  • Page 1 8-cell truck scale that is being used with two iQUBE CPU boards. Load cells or simulators must be connected before using the instructions in this document. See the iQube2 installation manual (available on the Rice Lake website at www.ricelake.com).
  • Page 2 COMM SEL PORTTYPE DIGIO COMM BOARDS CELLS SCALES iQube Boards Number of iQube2 boards on port Save Manual Auto Assign Assign and Exit Figure 3. BOARDS Parameter 2. Press and release the button on secondary board #1. Cells 5-8 are now available to the system SETUP (S1) and registers its serial number with the primary board.
  • Page 3 12/20/2010 01:57PM Primary 123XX456 Secondary 456XX789 _________ _________ Auto Assign Done Edit Figure 4. Primary and Secondary Boards screen 5. Press the softkey. Done Configure and Add Load Cells 1. Under , set the capacity and m/v. CELLS 2. Add cells 5-8 to Scale 1 under the »...
  • Page 4 Port General iQube2 type to RS-485 Configure and Add Load Cells and Secondary Board 1. Select the iQUBE tab. Figure 7. Serial port iQUBE2 settings 2. Press and update as needed. Load Cell Configuration iQUBE Quick-Start Guide...
  • Page 5 Save As 2. From the menu, select Communications Connect 3. Download all configuration steps. Mark the checkbox. Download to iQube2 4. On the , press the softkey. A weight displays. 920i Save and Exit 5. Continue with calibration using installation manual.
  • Page 6 Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319...

Table of Contents