Rice Lake iQUBE2 Enterprise 1280 Installation Manual
Rice Lake iQUBE2 Enterprise 1280 Installation Manual

Rice Lake iQUBE2 Enterprise 1280 Installation Manual

Digital diagnostic junction box, indicator, programmable hmi indicator, pc-based hme
Table of Contents

Advertisement

Quick Links

iQUBE2
®
Digital Diagnostic Junction Box
Version 1.0
Installation Manual
1280 Enterprise
Indicator
920i
Programmable HMI Indicator
®
Virtui2
PC-based HME
®
PN 106113 Rev D
December 20, 2019

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the iQUBE2 Enterprise 1280 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Rice Lake iQUBE2 Enterprise 1280

  • Page 1 iQUBE2 ® Digital Diagnostic Junction Box Version 1.0 Installation Manual 1280 Enterprise Indicator ™ 920i Programmable HMI Indicator ® Virtui2 PC-based HME ® PN 106113 Rev D December 20, 2019...
  • Page 2 All information contained within this publication is, to the best of our knowledge, complete and accurate at the time of publication. Rice Lake Weighing Systems reserves the right to make changes to the technology, features, specifications and design of the equipment without notice.
  • Page 3: Table Of Contents

    2.12 Parts Kit Contents ................. . . 20 Technical training seminars are available through Rice Lake Weighing Systems.
  • Page 4 Temporary Zero Calibration ..............54 Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost.
  • Page 5 Specifications ..................78 Technical training seminars are available through Rice Lake Weighing Systems.
  • Page 6 Digital Diagnostic Junction Box Rice Lake continually offers web-based video training on a growing selection of product-related topics at no cost. Visit www.ricelake.com/webinars Visit our website www.RiceLake.com...
  • Page 7: Introduction

    Do not operate or work on this equipment unless this manual has been read and all instructions are understood. Failure to follow the instructions or heed the warnings could result in injury or death. Contact any Rice Lake Weighing Systems dealer for replacement manuals.
  • Page 8: Overview

    The CPU board (PN 106390) provides four load cell input channels, four configurable I/O channels, a selectable multi-protocol communications port (RS-232, 2- or 4-wire RS-485) and an additional port that accepts various Rice Lake option cards (see Table 1-2 on page 3) for simultaneous communications with multiple hosts.
  • Page 9: System Configurations And Options

    Supporting Indicators The iQUBE2 supports any indicator as a serial scale. The 1280 and 920i provide special functionality to work in conjunction with the iQUBE2, allowing for configuration, calibration, and diagnostics. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 10: Printed Information

    iQUBE2 Digital Diagnostic Junction Box Printed Information iQUBE2 Options • Fiber-Optic Interface Installation Instructions, PN 80230 • Ethernet Communications Installation Manual, PN 77224 • USB Communications Card Installation Instructions, PN 95490 • Serial Communications Expansion Card Installation Instructions, PN 110892 •...
  • Page 11: Installation

    Immediately after unpacking, visually inspect the iQUBE2 and the parts list to ensure all parts are included and undamaged. If any parts were damaged in shipment, notify Rice Lake Weighing Systems and the shipper immediately. Parts kit contents for 4- and 8-channel models of the iQUBE2.
  • Page 12: Cable Connections

    iQUBE2 Digital Diagnostic Junction Box Cable Connections All models of the iQUBE2 provide cord grips for cabling into the unit, plus a dedicated cord grip for a ground wire. Up to 12 load cells can be cabled into the iQUBE2 (in large FRP model only); other cord grips allow cabling for serial communications, digital I/O, and AC power.
  • Page 13: Transient Protection

    12 gauge wire recommended Power Company Ground Rod NOTE: Always verify that installed wiring is properly grounded Figure 2-3. Truck Scale Single Point Grounding Example © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 14: Load Cells

    Diagnostics may not function properly with load cells having a bridge resistance greater than 700 ohms. IMPORTANT Customers wanting to use load cells with a higher bridge resistance than 700 ohms can consult the Rice Lake Weighing Systems service department for additional options.
  • Page 15 HB1 is a heartbeat monitor which monitors processor activity. Blinking once per second indicates normal operation. Note As either errors or COM port activity increases, HB1’s blinking may slow down to reflect the added activity. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 16: Serial Communications

    iQUBE2 Digital Diagnostic Junction Box Serial Communications Serial communications between the iQUBE2 and host or peripheral devices is provided by two ports. • Port 1 – provides onboard RS-232 (full-duplex) or RS-485 (full- or half-duplex) communications on connector J7. S2 on the CPU board is a pair of switches that are used to select the required setting •...
  • Page 17: Communications Option Cards

    48 Mbps 36 Mbps 115' 24 Mbps 140' 18 Mbps 170' 11 Mbps 180' 9 Mbps 190' 5.5 Mbps 215' 3 Mbps 260' 1 Mbps 330' Table 2-7. Approximate 802.11g Data Rate/Range © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 18: Digital I/O

    If preacts are required, the host device should be used as the controller for its own I/O. Rice Lake does not recommend the use of iQUBE2 (local) I/O with host device controlled setpoints using the preact functions as variances in the reaction time in serially controlled I/O can dramatically affect the preact algorithm.
  • Page 19: Primary/Secondary Wiring And Configuration

    J1–J4 of the first secondary unit assigned cells 5–8, as shown in Table 2-11 Load Cell Primary Cell Secondary #1 Cell Secondary #2 Cell Secondary #3 Cell Connector Numbers Numbers Numbers Numbers Table 2-11. Cell Numbering for Multiple-Board System © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 20: Assigning Board Addresses

    iQUBE2 Digital Diagnostic Junction Box 2.8.1 Assigning Board Addresses Once the primary and secondary unit ports are wired (Figure 2-4 on page 9), their addresses must be assigned. There are two ways to assign the secondary board addresses in an iQUBE2 system: manual entry or automatic, push-button assignment. 2.8.2 Manual Entry Manual address assignments are done by manually entering the secondary board IDs and the desired addresses in the Virtui2...
  • Page 21: Restore System Defaults

    3. Set SW1 to OFF to overwrite stored system parameters with default values. Power cycling the unit without overwriting stored parameters reloads stored parameters including previously Note assigned secondary board IDs and addresses. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 22: Installing Option Cards

    iQUBE2 Digital Diagnostic Junction Box Installing Option Cards Each option card is shipped with installation instructions specific to card. Follow general procedures for installation into iQUBE2. Option cards are not hot-pluggable. Disconnect power to the iQUBE2 before installing option cards. CAUTION 1.
  • Page 23 Board (J6) power + VDC Figure 2-10. Transient Protection Board Wiring 9. Replace the enclosure cover. 10. Reconnect the power to the iQUBE2. See Addendum (PN 160051) for initial installation instructions. Note © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 24: 2.11 Replacement Parts

    iQUBE2 Digital Diagnostic Junction Box 2.11 Replacement Parts Brown Wire Top View - Cover Removed Top View Figure 2-11. iQUBE2 Parts Illustration www.ricelake.com Visit our website...
  • Page 25 Main power supply fuse, 2.5 amp 5x20mm 80869 CPU board fuse 1122262 Refresh cable Replace fuses only with fuses of the same type and rating to protect against CAUTION the risk of fire. Table 2-13. Replacement Parts © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 26: 2.12 Parts Kit Contents

    iQUBE2 Digital Diagnostic Junction Box 2.12 Parts Kit Contents Parts kit contents for the 4- and 8-channel models of the iQUBE2. Quantity Description 4-ch 8-ch 14861 Machine screws, 8-32NC x 3/8 15134 Lock washers, No. 8 Type A 15631 Cable ties, 3-in. 15665 Reducing glands, 1/2 NPT 19538...
  • Page 27: 1280 Configuration

    Figure 3-2. 1280 Scale Kind Settings can be made to the iQUBE2 prior to setting the scale kind; however, the 1280 must be set to the iQUBE2 Note scale kind before it can be used. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 28: Iqube2 Setup

    iQUBE2 Digital Diagnostic Junction Box 3.1.1 iQUBE2 Setup Popup – enter iQUBE2 setup Select iQUBE2 Figure 3-3. 1280 Communication Menu 1. From the Communication Menu on the 1280 select iQUBE2. A popup displays, asking to enter iQUBE2 setup mode. 2. Press to enter setup.
  • Page 29: Base Setup

    Digital I/O – select digital I/O and set parameters (Section 3.2.6 on page 26) Diagnostics – select diagnostics to diagnose various parameters (Section 3.2.7 on page 27) Table 3-2. Base Parameters © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 30: Board Assignments

    iQUBE2 Digital Diagnostic Junction Box 3.2.1 Board Assignments Set the board IDs and addresses. 1. Select Board Assignment. 2. Select the board to assign. A keypad displays. 3. Enter the appropriate ID or address on the displayed keypad. 4. Press after each parameter is set.
  • Page 31: Load Cell Assignments

    RS485 Address Command Echo Specifies whether characters received by port are echoed back to sending unit Streaming Checksum Streaming Prefix Streaming Postfix Polling Checksum Polling Prefix Polling Postfix Table 3-6. Communication Parameters © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 32: Digital I/O

    iQUBE2 Digital Diagnostic Junction Box To set the communication ports: 1. Select Comm Ports. 2. Select a comm port from the drop down list. 3. Select each parameter and enter the required information. 4. Press after each parameter is set. 3.2.6 Digital I/O The Digital I/O menu is used to assign functions to digital inputs and outputs.
  • Page 33: Diagnostics

    Peak to Peak Load Cell Noise Sets peak to peak noise threshold for all scales (in   Cell Noise Time Sets peak to peak noise timer (seconds) Off or on Cell Balance Diagnostics Table 3-8. Diagnostics Descriptions © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 34: Scales

    iQUBE2 Digital Diagnostic Junction Box Function Description Cell Balance Range Sets load cell balance range (% of capacity) Cell Balance Threshold Sets load cell balance threshold (% of capacity) Off or on Cell Drift Diagnostics Cell Drift Theshold Sets load cell drift threshold (% of capacity) Cell Drift Range Sets load cell drift range (% of capacity) Cell Drift Time...
  • Page 35: Filtering

    Display Divisions 1, 2, 3 Table 3-11. Formatting Parameters To select the format, 1. Select Formatting. 2. Select Primary or Secondary from the drop down menu. 3. Select Decimal Position © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 36: Communications

    iQUBE2 Digital Diagnostic Junction Box 3.3.3 Communications Parameter Default Description Gross Print Format FMT1 FMT1 through FMT6 Net Print Format FMT2 FMT1 through FMT6 Print Destination COM1 Off, COM Port 1, 2, 3, 1,2 1,3, 2,3 1,2,3 Stream Format FMT4 FMT1 through FMT6 Stream Destination Off, COM Port 1, 2, 3, 1,2 1,3, 2,3 1,2,3...
  • Page 37: Setpoints

    Performs functions based on the gross weight; the target weight entered is considered a positive gross weight Performs functions based on the net weight; the target weight entered is considered a positive net weight value Table 3-15. Setpoint Kinds © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 38: Calibration

    iQUBE2 Digital Diagnostic Junction Box Calibration Calibration types depend on how many load cells are connected to iQUBE2. If only one load cell is connected, only 2-point calibration is available. The iQUBE2 can be calibrated by using the Virtui2 Configuration Utility host device front panel, or by serial commands.
  • Page 39: 920I Configuration

    Save and Exit writes all parameter changes to flash memory before returning to weigh mode in both iQUBE2 and Note 920i. After parameter changes are complete, the configuration must be downloaded. See Section 4.2 on page © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 40: Connect To 920I

    iQUBE2 Digital Diagnostic Junction Box 4.1.1 Connect to 920i 1. From the SERIAL menu, navigate to the PORTTYPE sub-menu and select the correct port type. 2. Navigate to the CONFIG menu and press the softkey. Connect 4.1.2 Manual Assignment with the 920i 1.
  • Page 41: 920I-Iqube2 Menu Configuration

    Select measurement rate of the analog-to-digital converter, in samples per second; lower sample rates provide greater signal noise immunity; 30HZ (default), 50HZ, 60HZ, 100HZ, 500HZ, 2.5HZ, 5HZ, 10HZ, 15HZ, 25HZ WARMUP The delayed response time (minutes) during warm-up; 1-10 minutes, 6 (default) Table 4-1. Serial Menu iQUBE2 Sub-menu © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 42 iQUBE2 Digital Diagnostic Junction Box Parameter Description DIGIO Specify the function of each of the digital I/O bits; BIT 1 - BIT 4; Use Revolution to configure parameters OFF indicates the bit is not configured ZERO, TARE, NT/GRS, UNITS, PRINT — provide the same functions as the five major front panel keys CLRCN reset the consecutive number CLRTAR clear the current tare for the active scale SETPNT designate the bit as a free-running setpoint within the iQUBE2...
  • Page 43 Rx- B Rx+ A Table 4-3. 4-Wire RS-485 Connections for 920i Host (Continued) RS-485 is an addressed protocol. Rice Lake Weighing Systems recommends using RS-422, if there is only one Note iQUBE2 system. See 920i setup for port type. CONFIG Menu 1.
  • Page 44: Set Comm Parameters And Download Configuration

    iQUBE2 Digital Diagnostic Junction Box Associating Digital I/O 1. Enter the DIGIO parameter to specify the function of each digital I/O bit. By default, Bits 1-4 are set to OFF. Higher number bits (secondary boards) are set to cell health by default and cannot be changed. Revolution III should be used to fully configure the parameters shown in Table 4-4.
  • Page 45: Associating A Scale

    DFSENS value. To disable filtering, set DFTHRH to none. Note © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 46: 920I Calibration

    iQUBE2 Digital Diagnostic Junction Box 920i Calibration The calibration types available depends on how many load cells are connected to iQUBE2. If only one load cell is connected, only 2-point calibration will be available. The iQUBE2 can be calibrated by using the Virtui2 Configuration Utility 920i, or serial commands.
  • Page 47: Front Panel Calibration

    8. Press to accept the value. Basic calibration is complete. ENTER 9. Press the softkey to return to weigh mode. To continue with linearization, see Section 4.3.2 on page Save and Exit © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 48 iQUBE2 Digital Diagnostic Junction Box Cal-Match Calibration (Corner, Section) Cal-Match calibration can be performed using a vehicle as the test weight; however, a smaller footprint and heavier weight value often yields better results. Mechanical characteristics of the scale may prevent Cal-Match from automatically computing load cell trim factors.
  • Page 49: Advanced Calibration Options

    DOWN ENTER 3. Ensure all weight is removed and press the softkey to re-zero the scale. Calibrate 4. Press , then press twice to return to the CALIB menu. ENTER © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 50: Virtui2 Configuration

    iQUBE2 Digital Diagnostic Junction Box Virtui2 Configuration This section describes configuration for the iQUBE2 using the Virtui2 Configuration Utility. Instructions in this section assume that the configuration application is installed on the PC, and that load cell and communications wiring (between multiple iQUBE2 boards, if applicable, and between the iQUBE2 and the PC) is complete.
  • Page 51: Device Configuration

    4. Set the address to a value other than 1 to activate the device. Each secondary device must have its own unique address between 2-254. 5. Use the tab to specify the sample rate. General To deactivate a secondary device, set the address to 1. Note Figure 5-3. Device Configuration Screens © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 52: Communications

    iQUBE2 Digital Diagnostic Junction Box 5.1.1 Communications In this section communications settings and stream formats are set for the alternate communications port. 1. Select Communications. 2. Select the tab. General 3. Using the drop-down menus, specify the baud rate, terminator, end of line delay, prefix character, data bits/parity, echo, RS-485 address, postfix character, and stop bits as desired.
  • Page 53: Load Cell Assignment

    Section Layout provides essential information to the system about cell relationships. Calibration and diagnostic functions depend on an accurate representation of the scale. Load cells do not have to be on the same board to be paired with each other. Note © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 54: Scale Configuration

    iQUBE2 Digital Diagnostic Junction Box 5.1.4 Scale Configuration Scale configuration can be accomplished via the System section in the left pane. Click System to expand this section. 1. From the left pane, with the System section expanded, click Scales. 2. From the listing of scales, click on the one to be configured. Figure 5-7.
  • Page 55: Sending Configuration To Iqube2

    (located at the bottom of the window) ’s Connect indicates a connection. Figure 5-9. Connection Icon 3. Once connected, click from the tool bar, or select iQUBE2 from the Send Configuration Send Configuration to Communications menu. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 56: Configuration Utility Calibration

    iQUBE2 Digital Diagnostic Junction Box Configuration Utility Calibration Calibration types depend on how many load cells are connected to iQUBE2. See Table 5-2. If only one load cell is connected, only 2-point calibration will be available. The iQUBE2 can be calibrated by using the Virtui2 Configuration Utility host device or serial commands.
  • Page 57: 2-Point Calibration

    10. If linear points were selected in Step 4, enter the applied weight in the text box and press . Success displays. Measure Figure 5-12. Two-Point Calibration Wizard 11. Press Next © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 58: Corner Match

    iQUBE2 Digital Diagnostic Junction Box 12. If chains or hooks are being used, remove them (along with all other weight) from the scale and select Re-Zero Calibrate. 13. Press . Calibration is complete. Next 14. Press to commit the calibration values and send them to iQUBE2. Finish Press Cancel to abort;...
  • Page 59: Section Match

    Success displays when the Measure measurement is finished. As each measurement is completed, the next cell becomes available to measure. 11. Repeat steps 9-10 until each section has been measured. 12. Press Next © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 60: Theoretical Calibration

    iQUBE2 Digital Diagnostic Junction Box 13. If certified test weights are not used, a span calibration must be completed. If they were used (Step 7), the calibration was completed during normalization. 14. Add the test weight to the scale and press Calibrate Span 15.
  • Page 61: Serial Commands

    Show primary board power supply voltage WHOAMI Returns information about iQUBE2 including master ID, version, port configuration. * These commands also store the configuration to EEPROM. Table 6-2. iQUBE2 System Commands © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 62 iQUBE2 Digital Diagnostic Junction Box Command s = scale number Description Start streaming for all scales End streaming for all scales SCs.SX Start streaming for scale s SCs.EX End streaming scale s Write one frame of polling format from all scales Write one frame of streamed format from all scales SCs.COUNTS Read A/D counts...
  • Page 63 Read calibrated weight value — LCx.SAMPLEn Uploads/Downloads calibration values for -16777215–16777215 the given (n) Cal-Match calibration step ( LCx.SAMPLE Uploads/Downloads calibration values for -16777215–16777215 all calibration steps Table 6-6. iQUBE2 Load Cell Commands © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 64 iQUBE2 Digital Diagnostic Junction Box Command = scale number Description Values SCs.CALTYPE Calibration type SECMATCH, CORNERMATCH, 2-POINT Calibration commands for CALTYPE = SECMATCH, CORNERMATCH SCs.CAL0 Zero calibration — SCs.CALn Calibration step n Assign calibration step number (1–16, maximum) for each load cell in CORNERMATCH calibration, for each section pair (1–8, maximum) in SECTION- MATCH calibration SCs.NORM...
  • Page 65 Set digital output n on DOFF#0 Set all digital outputs off (inactive) — 1–4 DOFF#n Set digital output n off MSDOUT Reads all outputs in the system — Table 6-10. iQUBE2 Digital I/O Commands © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 66 iQUBE2 Digital Diagnostic Junction Box Section 6.0 on page 55 for more information about the commands listed in Table 6-11. Command = scale number Description Values DIA.PS Query power supply error — DIA.MSCONNECT Query secondary connection error — DIA.PSEXC Query excitation error —...
  • Page 67: Diagnostics

    Cell connection DIA.CELLCONNECT Zero reference DIA.ZREF Overloaded cell DIA.OVERLOAD Drift DIA.DRIFT Underloaded cell DIA.UNDERLOAD Peak noise DIA.NOISE Unbalanced cell DIA.UNBAL <space> No error — Table 7-2. Diagnostic Codes and Query Commands © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 68: Retrieving Diagnostic Data

    iQUBE2 Digital Diagnostic Junction Box 7.1.2 Retrieving Diagnostic Data Table 7-2 also lists the diagnostic query commands associated with each of the diagnostic codes. When a diagnostic code is generated, the related query command can be used to retrieve additional information about the error. Using the query command for a given error also clears the diagnostic flag for that error.
  • Page 69: System Diagnostic Commands

    For load cells with a resistance greater than 2,200 Ohms, a high precision 2k Ohm resistor should be placed between Note the EXC+ and EXC- pins of the load cell connection to maintain correct operation of the Cell Connect diagnostic. Precision resistor properties: Temperature Coefficient ±10 ppm/°C Tolerance ±1% © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 70: Scale Diagnostic Commands

    iQUBE2 Digital Diagnostic Junction Box Cell Overload Error, <D> = V The overloaded cell diagnostic flag is set when the cell output exceeds a specified percentage of full scale cell output set for the scale. Query Command Example: DIA.OVERLOAD DIA.OVERLOAD=SC1 100.0% 4:15.233; The cell overload query, DIA.OVERLOAD, returns information including the affected scale number (SC1), the cell overload threshold (100% of full-scale output), the affected cell number (4:), and the millivolts value of the overload error (15.233 mV).
  • Page 71 5% of capacity.) The default value is 2 (2%); allowed values are 0–100. SC1.DIA.DRIFT.TIME=15 Sets the number of seconds to wait between successive cell drift diagnostic tests to 15 seconds. The default value is 10 (10 seconds); allowed values are 0–300 seconds. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 72 iQUBE2 Digital Diagnostic Junction Box Cell Noise Error, <D> = N The noise diagnostic flag is set when a single cell experiences noise greater than the configured noise threshold. Query Command Example: DIA.NOISE DIA.NOISE=SC1 1.00 1:1.10 3:1.08; The cell noise query above, DIA.NOISE, returns information including the scale number (SC1), the value set as the system noise threshold (1.00 V), and the noise values of two cells (1: and 3:) that exceed the threshold (1.10 and 1.08 V).
  • Page 73: Diagnostic Flag Commands

    Table 7-4. Mask Bits and Codes for Diagnostic Errors 0x100–0x800 0x010–0x080 0x001–0x008 — — Table 7-5. Diagnostic Flag Mask Bits The DIA.CLEAR command clears all diagnostic flags. The completed command responds with an OK. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 74: Data Formatting

    iQUBE2 Digital Diagnostic Junction Box Data Formatting Weight Data Formats The following data formats are available for the iQUBE2. Each of the listed formats can be customized to suit the application. Identifier Default Format FMT1 (default gross format) Scale<SC>:<CR><LF>Gross<TAB><G><CR><LF> FMT2 (default net format) Scale<SC>:<CR><LF>Gross<TAB><G><CR><LF>Net<TAB><N><CR><LF>...
  • Page 75: Weight Data Status Codes

    Field Status Code Meaning <D> Power supply Secondary communication Excitation Cell connection Zero reference Overloaded cell Drift Underloaded cell Peak noise Unbalanced cell <space> No error Table 8-4. Diagnostic Status Codes © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 76: Data Formatting Examples

    iQUBE2 Digital Diagnostic Junction Box Data Formatting Examples 8.2.1 Formatting Data from Multiple Scales Data from multiple scales can be streamed or polled using one or several formats, to one or more comm ports. Example, given a four-scale system with the following data format assigned to FMT1: FMT1=<SC><P><W7.><S>...
  • Page 77: Maintenance And Troubleshooting

    Junction Box). On multi-cell systems, emulation performance is highly dependent on the ability to center the load with respect to all the load cells in the system. In general, the closer the load footprint size is to the size of the deck, the better the emulation performance will be. © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 78: Cell Emulation In Virtui2

    iQUBE2 Digital Diagnostic Junction Box 9.1.4 Cell Emulation in Virtui2 To enable cell emulation in Virtui2: 1. Log in to the Virtui2 Configuration Utility as a user with administrator rights (see Section 5.2 on page 50). 2. If the bad load cell needs to be identified, use the in the menu to make that Diagnostics Monitor...
  • Page 79: System Current Draw

    Transient Board Current Load Voltage Drop 1.0 A 0.16 V 1.5 A 0.24 V 2.0 A 0.32 V 2.5 A 0.40 V 3.0 A 0.48V Table 9-3. Transient Board Voltage Drop © Rice Lake Weighing Systems ● All Rights Reserved...
  • Page 80 iQUBE2 Digital Diagnostic Junction Box Power Cable Voltage Drop Power cable voltage drop is a function of the type of cable used and the length of the cable run. The run length includes both the supply and return lengths. Stranded Wire Gauge ...
  • Page 81: Installing Firmware Updates

    The iQUBE2 Flash Updater application is automatically installed with Virtui2. If using Virtui2, the iQUBE2 Flash Updater can be run by clicking iQUBE2 Start » All Programs » Rice Lake Weighing Systems » Flash Updater Because the iQUBE2 Flash Updater uses the same port as Virtui2, the Virtui2 services must be stopped before the Note iQUBE2 Flash Updater can execute a firmware update.
  • Page 82: Replacing Cpu Boards

    iQUBE2 Digital Diagnostic Junction Box iQUBE2 J14 Connections DB-9 Female Connector Signal Signal Table 9-5. Host Communications DB9 Connections (Continued) 2. In the iQUBE2 Updater, use the pull-down menu to select the communications port of the PC that the COM Port iQUBE2 is connected to.
  • Page 83: Troubleshooting

    If the configuration looks good, check the outputs of each load cell. Cell outputs should change when a weight on top of each cell or when the load cell cable is disconnected. If these steps do not resolve the problem, contact Rice Lake Weighing Systems technical support for further assistance. 9.5.2...
  • Page 84: Specifications

    iQUBE2 Digital Diagnostic Junction Box Specifications AC Model Power Serial Communications Line Voltages 115 or 230 VAC Port 1 RS-232/422/485 Frequency 50 or 60 Hz 9,600 - 460,800 Baud Current Draw 2 A (max per 4 CPU system) Port 2 Options: Fiber-Optic, Ethernet, DC Model Power Wireless Ethernet, USB,...
  • Page 86 Specifications subject to change without notice. Rice Lake Weighing Systems is an ISO 9001 registered company. 230 W. Coleman St. • Rice Lake, WI 54868 • USA U.S. 800-472-6703 • Canada/Mexico 800-321-6703 • International 715-234-9171 • Europe +31 (0)26 472 1319...

This manual is also suitable for: