DentalEZ® NuSimplicity™ Chair. the features, operation procedures, and care guidelines with the doctor’s staff . The NuSimplicity Chair is intended to be used by trained professional dental care personnel only. Leave this manual in the doctor’s off ice.
Superior ergonomic design minimizes twisting and reaching. In order to achieve the most eff icient use in the operatory and understand what the NuSimplicity Chair off ers, here is a compilation of some of those features and benefi ts.
Section I Introduction NuSimplicity Chair Range of Motion Figure 3. Range of motion for the NuSimplicity dental chair www.DentalEZ.com 866-DTE-INFO PN: 2717-256E...
Section I Introduction NuSimplicity Chair Classifi cations Explanation of Symbols & Signs Medical-General Medical Equipment = Caution Certifi ed as to electrical shock, fi re and mechanical hazards only in = Warning 53HN accordance with: RISK UL 60601-1 CLASS = Biohazard CAN/CSA C22.2 NO.
• Rating of main circuit breakers should be 20 Amps maximum. • Do not modify the NuSimplicity Chair without permission from DentalEZ. • To satisfy FCC RF exposure requirements for mobile and base station transmission •...
Page 11
Section I Introduction NuSimplicity Chair Safety Precautions (Continued) NOTICE NOTICE • Medical electrical equipment needs special This equipment has been tested and found precautions regarding electromagnetic to comply with the limits for a Class B (EMC) compatibility and needs to be digital device, pursuant to Part 15 of the FCC installed according to EMC information.
1. Using pliers, remove the staples that secure the shipping carton to the wooden pallet. • The NuSimplicity Chair is shipped with a retaining strap to secure the base 2. Remove the carton and packing inserts from mechanism.
To prevent injury as a result of chair chair, remove and set aside the following items tipping, chair must be placed on a smooth, from the NuSimplicity Chair upholstery assembly hard and level fl oor. packaging (Figure 4): • DO NOT position the chair any place •...
Cover 1. Remove the following hardware from the bag of supplies: Spring Cover Bracket Figure 6. NuSimplicity Chair spring cover • 10-32 Pan Head Screws (4) • Spring Cover Brackets (2) • Spring Cover Bracket Screws (8) 2. Install a spring cover bracket (Figure 6) to the seat frame tube.
Chair Delivery Units Wired Touchpad Control Install console mounted delivery units and The NuSimplicity Chair’s primary touchpad Magellan style delivery units according to the controllers are located on both sides of the chair’s manufacturer’s instructions supplied with the back. One wire-type controller can be added and delivery unit.
Wireless Controls Tools Required If wireless foot control and/or wireless touchpad control options were ordered, they must be paired • Phillips-head Screwdriver with the NuSimplicity Chair. WARNING Tools Required To prevent any chance of electrical shock, • Phillips-head Screwdriver always disconnect power when indicated.
Page 17
Section III Installation NuSimplicity Chair Wireless Controls (Continued) 7. Chair can have up to four (4) wireless Pairing Switch Battery Compartment connections. Using chair back touchpad control, press and hold 1, 2, 3 or A button until chair double beeps to store appropriate memory location (Figure 14).
For proper air glide operation, the fl oor MUST BE LEVEL and have a smooth, hard and non- obstructive surface. 1. Place the NuSimplicity Chair on a smooth, hard and level (vinyl, tile or terrazzo type) fl oor. Figure 16. Air regulator toggle switch 2.
Section III Installation NuSimplicity Chair Upholstery Back (Factory Installed) Seat NOTE: Make sure the blind grip fasteners are 1. Obtain the 1/4-20 thumb screw provided in the fi rmly secured to the back of the upholstered hardware package. chair back.
Page 20
Section III Installation DentalEZ Equipment ® Upholstery (Continued) Headrest 2. Place the magnetic headrest cushion on the chair or the fi xed headrest of the chair. 1. Slide the blade of the standard fi xed headrest (Figure 20) or the optional double articulating...
Section III Installation NuSimplicity Chair Installation Checklist NOTICE Before performing checklist task, make sure the area is at room temperature. Failure to comply can result in a faulty operation. With someone sitting on the chair, run each function a few times to ensure proper function.
Base Lowering Safety Switch Chair Rotation The NuSimplicity Chair is equipped with a base The NuSimplicity Chair is capable of swiveling 30° to lowering safety switch. This feature ensures either side of its center position. nothing gets caught underneath the chair base as To swivel the chair: it is lowered.
NuSimplicity Chair Armrests Standard Fixed Headrest Both of the NuSimplicity Chair’s armrests are To position the standard fi xed headrest: designed to be lowered, allowing easy patient Firmly push down or pull up the blade from the entry and exit from either side of the chair.
Section IV Operation DentalEZ Equipment ® Double Articulating Headrest (Opt) Air Glide (Optional) To position the optional double articulating The optional air glide aids the operator in headrest: repositioning the chair. 1. While facing the back of the headrest, fully To operate the air glide option: depress the release mechanism (Figure 26).
To prevent possible injury caused by accidental operation, do not leave young children unattended. Manual positioning of the NuSimplicity Chair is accomplished by using the touchpad controls located on each side of the chair back, the foot control, or the delivery unit touchpad control (Figure 28).
NOTE: Pressing any button on a chair control NuSimplicityChairSettingsVideo module while the NuSimplicity Chair is moving to a programmed position will immediately stop the movement of the chair, and three short beeps will Programming Auto Positions sound. To continue and complete the automatic...
Section IV Operation NuSimplicity Chair Programming Chair Travel Limits An LED display is located on the upper cantilever cover of the chair (Figure 29). This display indicates Power, Service and Run modes. Upper Limits LED Display 1. Press 1 and 2 on the chair mounted touchpad for fi ve (5) seconds.
Section IV Operation DentalEZ Equipment ® Control Valve Speed Adjustment b. Move the back up and down, adjusting the fl ow WARNING controls until it takes 12 seconds for the back to move either up or down the full travel.
• Using a soft cloth dampened in clean water, wipe away any residue and dry. Before attempting to clean or disinfect the NuSimplicity Chair, please read the following 2. To remove heavy soil: instructions carefully. • Dampen a soft, white cloth with lighter Since any cleaning product may be harmful or fl uid (naphtha) and rub the area gently.
Page 31
Section V Care NuSimplicity Chair Cleaning (Continued) Plastic and Coated Metal Surfaces • Allow the bleach solution to puddle on the aff ected area, or apply a bleach solution- soaked cloth to the area for approximately CAUTION 30 minutes. • NEVER use abrasives or petroleum-based •...
Metrex Research of material from the following manufacturers: Cavicide is recommended for disinfecting all other surfaces of the NuSimplicity Chair. The use of MorCare Collection from Morbern any disinfectant other than Cavicide may cause MorGuard from Morbern was developed to...
® Service Instruction Disposal of Equipment Disposal and Decommissioning of If the area of concern is not addressed in this manual, contact your local DentalEZ full-service DentalEZ Products dealership at 866-DTE-INFO (866-383-4636). See NOTE: All local regulatory requirements for Limited Warranty.
Section VII Parts Lists/Diagrams DentalEZ Equipment ® Hydraulic Controls Part/Kit Name Part/Kit No. Part/Kit Name Part/Kit No. Pump (115V AC) 3802-451 Touchpad Switch, RH 3802-423 Pump (220V AC) 3802-452 Touchpad Switch, LH 3802-441 Valve (115V AC) 3802-434 Touchpad Circuit Board...
Page 37
Section VII Parts Lists/Diagrams NuSimplicity Chair DentalEZ Parts Online To order parts online, visit www.dentalezparts.com or scan QR code with smartphone. www.DentalEZ.com 866-DTE-INFO PN: 2717-256E...
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The NuSimplicity Chair is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the NuSimplicity Chair should assure that it is used in such an environment.
Page 39
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions The NuSimplicity Chair is intended for use in the electromagnetic environment specifi ed below. The customer or the user of the NuSimplicity Chair should assure that it is used in such an environment.
NuSimplicity Chair DentalEZ and its employees are proud of the products we provide in the dental community. We stand behind these products with a warranty against defects in material and workmanship as provided below. In the event you experience diff iculty with the application or operation of any of our products, please contact our Technical Service Department at our expense at 866-DTE-INFO (866-383-4636).