Thank you for purchasing the Soehnle Professional Terminal 3025. This terminal was developed according to the requirements of operational practice and offers you numerous possibilities to individually adapt the weighing process to your needs in order to make your processes rational and cost-effective.
TABLE OF CONTENTS PAGE 1. GENERAL INFORMATION ......................6 1.1 Purpose of use ............................. 6 1.2 Safety instructions..........................6 1.3 Cleaning ............................... 6 1.4 Maintenance and service ........................7 1.5 Warranty/Liability ..........................7 1.6 Installation instructions ........................8 2. DEVICE DESCRIPTION ........................ 9 2.1 Connection and device description ....................
Page 4
6.3 Navigation in the application programs ................... 58 6.4 Special programs..........................58 6.5 Weighing and Taring ......................... 59 6.5.1 Function keys ..........................59 6.5.2 Display view ..........................59 6.5.3 Display information field ......................60 6.5.4 Weighing without taring ......................61 6.5.5 Manual taring ..........................
Page 5
6.8 Checking ............................99 6.8.1 Function keys ..........................99 6.8.2 Display view ..........................100 6.8.3 Display information field ......................101 6.8.4 Tare function ..........................101 6.8.5 Control function ........................101 6.8.6 Totalising and Picking ......................104 6.8.7 Display options with the Info key .................... 108 6.8.8 Transfer of current values to the fixed memory ..............
1. GENERAL INFORMATION 1.1 Purpose of use The Soehnle Professional Terminal 3025 is a calibratable terminal for use in scales and weighing systems with standard DMS weighing and load sensors. Up to two analogue measuring points (scales) can be connected internally at the same time. Another scale can be connected via the serial interface, e.g. as a reference scale.
Insofar as Soehnle Industrial Solutions provides a guarantee in individual cases, this means that the delivery item is free from defects during the guarantee period. Please keep the original packaging for any return transport.
1.6 Installation instructions Please check packing and terminal for external damage. Do not operate a visibly damaged terminal, but inform your dealer. Please keep the packaging for possible shipment. The terminal is designed for rough commercial use. However, it is a sensitive instrument whose performance can be affected by unfavourable environmental factors.
Colour display TFT 3.5" QVGA, dot matrix 320 x 240 pixels dimmable backlit > Option: Soehnle Professional Micro-SD card (high-quality SLC memory card), memory capacity: fixed memory for a total of 999 fixed values across programs, alibi memory for 4 million entries and for software updates >...
2.3 Symbols in the display Symbols in the display toolbar Alibi memory Alibi memory Alibi memory fill Key lock Data logger fill level 100% fill level 80% level empty Multi range 1 Multi range 2 Multi range 3 Resolution X10 Zero Hold function active Symbols in the information field of the display...
The scope of this information is Germany. In other countries, conformity with national laws must be checked. Terminal 3025 is approved within the EU for Class III and IIII custody transfer. It corresponds to the type described in the type approval as well as to the applicable requirements of Directives 2014/31/EU, 2014/30/EU, 2014/35/EU and EN 45501.
Switch on: Press the ON/OFF button When switching on, the logo, terminal type 3025, filter type and the detected measuring points are briefly displayed during the start routine. The scale is set to zero after power-up. When the scale is switched on, the last application program used is reactivated.
3. OPERATING CONCEPT The operating concept is divided into two functional areas: Setting mode • Program mode • 3.1 Setting mode Here you can make individual settings on the terminal to optimize the scale for your requirements. Table setting mode: Setting mode Program settings Terminal...
3.2 Program mode Select the desired weighing application program here. Once you have selected the application program and called it up, the scale is ready to weigh. 3.3 Function keys The function keys F1 to F6 are available for calling the functions. These assume different functions depending on the program and situation.
Page 15
Info key, electronic nameplate with approval data, Calling information about the application programs Function change key Scale change key Tare key Zero setting key / delete tare Enter key Print key On / Off Display: The display shows different views depending on the situation. It is divided into the following display fields: Symbol bar SYMBOLLEISTE...
4. SETTING MODE The setting mode stores program and weighing parameters that can be called up and individually configured. Adjustment can be made by entering data or selecting from predefined parameters. The individual parameters are stored in logical groups within a multi-level family tree structure. Note: The modification options may be limited in the legal metrology sector.
4.1 Calling up the setting mode Switch on the terminal. After the power-on routine, the terminal will automatically start the application program that was last activated. To call up the setting mode, use the function change key to switch to the setting functions level.
The Service area is protected by a password. The contained parameters can only be maintained by trained service personnel. Please contact your Soehnle Industrial Solutions service center. 4.2 Program settings In the Program Settings setting mode, you can adapt the user programs of your scale to your specific requirements.
> green > white > light blue > yellow > blue > magenta > red Factory setting: white Language Selection 1. language German (factory setting) 2. language English or other languages can be customized via the Soehnle Professional service program.
Page 20
Display Auxiliary display Selection > Second unit > Tare value > x10 Resolution > Gross/net > Current weight > Reference quantity > Reference weight > Setpoint > Difference to setpoint > Signal light Factory setting: Second unit Decimal separator Selection Comma, dot or none.
Page 21
Date / time Time Input via numeric keypad, clock starts to run Press the Enter key Date Input via numeric keypad Date format Selection European or imperial date Factory setting: European Input via numeric keypad in days [d] Use-by period (Days) Factory setting: 0 d Organisation data Description ORG A...
Page 22
Organisation data ORG data A fixed values Selection of existing fixed values from the database Scroll with the arrow keys F3 and F4. Call the desired fixed value with F6 Edit. Editing an existing fixed value from the database Press F6 to call up editing of the selected fixed value. Parameter number is marked, call via F6 Edit, input via numeric/alpha keyboard.
Page 23
Error memory Error list Request the error list with position, date, time of last error, number of errors, error number and description. Scroll back using the arrow keys or F5. Protocol memory Query the list of the download protocol memory with position, date, time, code no. and description.
Page 24
SD card data backup Import By accepting F6 Next the current user settings are loaded from the micro SD card. The security prompt Import comes with YES or NO. If NO is selected, no data is loaded. Export The current user settings are saved on the micro SD card by accepting F6 Continue.
4.4 Setting mode – Scale settings In the first step, select the scale (measuring point) with the scale change key, for which you want to make settings. Only connected and active scales (measuring points) are displayed for selection. The settings and queries are made individually for each scale (measuring point). All setting and query options are identical for all scales (measuring points).
Page 26
Scale parameters Zero range at power-on Query off or on When the scale is turned on, it sets the zero range of the weighing range to zero when the power is turned on. Selection only possible in the service area, subject to calibration. Auto tare Selection off or on off: No auto tare...
Page 27
Query of the weight value of the 4th support point weight value Reference point 4 Query of the adjusted raw measured value from the 4th support point raw measured value The internal measuring point is located on the central board of terminal 3025 with measuring point number 01.
Internal measuring point Operating mode Query or selection off or on With off the internal measuring point is deactivated with on measuring point 01 is active. Factory setting: on 4.5 Setting mode – Communication settings In the setting mode under Communication you will find the following query and setting options Alibi memory Memory capacity Query...
Page 30
Interface Port 2 (Further details see also interface description 470.508.090) Data bits Selection > 7 > 8 Factory settings: 8 Parity Selection > none > even > odd Factory settings: none Xon/Xoff Selection off or on Factory settings: off Factory settings Selection Do not execute: Individual settings on the interface are retained Execute: Resets interface to factory setting...
Page 31
Interface Port 3 (Further details see also interface description 470.508.090) Xon/Xoff Selection off or on Factory setting: off Factory settings Selection Do not execute: Individual settings on the interface are retained Execute: Resets interface to factory setting (Further details see also interface description 470.508.090) USB mode Selection >...
Page 32
Ethernet (Further details see also interface description 470.508.090) Deactivate MAC Address- Selection off or on bypass Automatic transmission of UDP-MAB message for recognition of our scale (for content, see User Mode 470.702.112) Factory setting: off Print template Request printout (OK) Selection on or off.
Page 33
Data processing settings (Further details see also interface description 470.508.090) Data processing mode Selection for EDP mode (terminal automatically triggers the selected data set) > no > A: Send data record 1 x immediately > B: Send data record 1 x at standstill + change >...
Page 34
Factory settings Selection Do not execute: Individual EDP settings are retained Execute: Resets EDP settings to factory settings Barcode (further details see also manual 3025 Chapter 7.9) Selection off or on Decimal separator numerical Factory setting: on Show barcode timer...
Page 35
Like filter for scan (ORGA) valid for filter for scan (ORGC) Filter for scan (ORGD) Like filter for scan (ORGA) valid for filter for scan (ORGD) Batch processing (further details see also manual 3025 Chapter 7.10) Batch code 1 Input Control commands from interface description 3025 (document 470.508.090D)
Page 36
Batch processing (further details see also manual 3025 Chapter 7.10) Batch code 4 parameter as Batch code 1 parameter source source Batch code 5 parameter as Batch code 1 parameter source source Batch activation key Selection > No key > Print key >...
4.6 Setting mode – Service settings The Service area is protected by a password. The contained parameters can only be maintained by trained service personnel. 4.7 Reset terminal To reset all settings (except scale parameters and service area) of the terminal back to the factory settings, you can perform a reset of the terminal with the selection in the Setting mode / Terminal / Reset Terminal and press the function key F6 Continue.
5. BASIC FUNCTIONS 5.1 Switch on The detailed description for switching on the terminal can be found in chapter 2.5 on page 12.. 5.2 Switch off The detailed description for switching off the terminal can be found in chapter 2.5 on page 12. 5.3 Program selection Selection from the available application programs.
Switch between several connected scales. Up to two analogue measuring points can be installed in the 3025 display unit. It is also possible to connect a digital scale via the RS232 interface with measuring point no. 30. Switching between the connected measuring points can be done manually or automatically.
5.8 Tenfold display x10 Displays the weight value with another decimal value in 10 times higher resolution. Use the Function Change key to switch to the Setting Functions level. Auxiliary Prog- Setting Gross x 10 display mode Press the key F2 x10. Approvable scale If the scale has been calibrated for legal metrology, the resolution is ten times higher as long as the x10 key is held down.
5.9 Hold function In setting mode, a hold function can be activated separately for each connected measuring point Setting mode / Scale / Scale parameters / Hold mode. The following hold functions are available for selection: > Not active: Hold is deactivated. >...
5.10 Auxiliary display Displays additional information, such as second units, weight values when counting, differences from the target value, etc. Activating and deactivating the auxiliary display Use the function change key to switch to the setting functions level. Auxiliary Prog- Setting Gross x 10...
Page 43
When you call the submenu Auxiliary display, the system displays a list of possible applications: You can move in the list with the arrow keys F3 and F4. Functions of the auxiliary display The following function states can be set in the submenu auxiliary display: >...
5.10.1 Second unit The second unit is defined in the setting mode for each connected scale. The following second units are stored as standard: Calibrated unit Second unit The following units are available: Unit Abbreviation Conversion to g Gram / 1 g / Kilogram 1.000 g...
Page 45
You can move in the list with the arrow keys F3 and F4. Press F6 "Apply" to define the second unit. After selecting the second unit, press F1 "End" to return to the weighing mode.
They are available in all application programs. The 3025 display unit has 4 organizational data memories. Each memory has > a freely selectable designation, e.g. "Article no.". (up to 12 characters, database up to 18 characters).
Page 47
ORG data terminals (ORGA to ORGD) from the working memory. Database for organizational data The use of the organizational data database is only possible with an optional Soehnle Professional Micro-SD card. Frequently recalled content can be stored as fixed values in a database and recalled as needed. This saves the repeated manual input of product names, operators, etc.
Page 48
For the requirement of a simple and multiple application to the care of the Org data base for the designations and contents there is the PC - user software 3025. The user software is connected to the PC via the EDP interface RS232, USB or Ethernet with the respective connection cable.
5.12 Alphanumeric input Alphanumeric entries can be made using the combined keyboard. Optionally, simplified input is possible with the USB interface and the USB adapter cable for PC keyboard 2550.03.027 via a USB - PC keyboard. Types of input fields Display fields without write capability Display/write fields with possibility of numerical input Display/write fields with alphanumeric input option...
5.13 Key lock Single, multiple or all keys can be locked on the keyboard. Lock keys Call up Terminal / Keypad / Keypad keylock in the setting mode. The keyboard is shown schematically in the display. By pressing the key, the status of the key can now be changed from open to locked and vice versa.
5.14 User password Access to the setting mode and program selection can be blocked by assigning an individual password to prevent unauthorized persons from making changes. Call up Terminal / User password in the setting mode. > Select New/Edit, in the input field that opens you can assign an alphanumeric password of up to 8 characters, then accept it with F6.
Page 52
F4 to select on and press F6 to accept. For the password request function, the login function setting must be activated. Without the internal password system in Terminal 3025, only an external EDP password is requested, sent and released.
Page 53
After switching on the terminal or the login call via the Enter key, the query window Login appears. With key F5 Cancel you get to the menu Security question Cancel login? Press F6 OK to display the password prompt. Enter the required password using the keyboard and confirm with the F6 key. The password with the command code is sent to the computer for verification and release and the hourglass appears in the display.
Page 54
Create/change password (internal terminal password) The Login function and Password query settings must be activated for the internal password query function. Call up the setting mode by pressing the F6 key in the Setting functions key level. Use the arrow key to highlight F4 Terminal, F6 Next. Use the arrow key to highlight F4 Login function, F6 Next.
Page 55
Enter the required password using the keyboard and confirm with the F6 key. The entered password is checked: If entered correctly, the computer record will be sent with the confirmation of registration and the terminal will then be in weighing mode. If the entry is incorrect, the display message Error 02 Incorrect password appears briefly and then the password query appears again.
5.16 SD - card data backup On the optional Micro SD card (Alibi memory) it is possible to save the setting parameters, to reset the data and functions at the terminal at a later time. Contents for data backup All setting parameters (user) and the program selection of the terminal are saved. The scale/adjustment parameters (ECAL) of the measuring points and the printouts are not saved.
6 APPLICATION PROGRAMS The various application programs of the Soehnle Terminal 3025 offer you comprehensive solutions for your weighing tasks. In the setting mode, you can adapt the application programs to your specific requirements. For more information, refer to the respective chapter Settings in Settings Mode.
6.3 Navigation in the application programs The function keys F1 to F6 are available for controlling the functions. They assume different functions depending on the program and situation. The respective function is shown on the display in the function bar. Depending on the program, the function keys are assigned in up to 4 levels, the change through the levels takes place rolling with the function change key.
6.5 Weighing and Taring Basic weighing function, determination of gross and net weights Note: The following description of the tare functions applies to the factory settings of the options > Auto tare in position off > Tare totalling in position, i.e. new tare overwrites old tare Selection of options in setting mode scale menu 6.5.1 Function keys The function keys F1 to F6 are available for controlling the functions.
Hint: To additionally display, for example, the tare weight or a second unit, you can activate an additional auxiliary display by pressing the F3 key. You can set the value to be displayed in the auxiliary display in the setting mode under Setting mode / Terminal / Display / Auxiliary display. Illustration: Auxiliary display with tare weight 6.5.3 Display information field In the Setting mode / Program settings / Weighing + taring / Display information field, you can set which data...
6.5.4 Weighing without taring Place the weighing piece on the scale, the weight is displayed with the message Gross. 6.5.5 Manual taring Determine and tare an unknown container weight. Place the empty container on the scale. Press the key The tare value is stored, the weight display goes to zero and the weight is displayed with the message Net. If the scale is unloaded, it displays the tare weight with a negative sign.
6.5.7 Multiplicative tare Taring several containers of the same weight with a known tare weight. Press the F4 Multi Tare key in the Tare functions level. Place a container on the scale or enter the known tare weight using the numeric keypad. Press the F6 key to confirm.
6.5.9 Provisional tare Unknown tare values are added to the existing tare memory; the net value remains unchanged. Press the F3 Prov. Tare key in the Tare functions level. Place the additional tare weight on the scale. Confirm with F6 Apply. The additional tare weight has been stored in the tare memory;...
6.5.11 Fixed tare memory Known tare weights can be stored in a power-fail-safe memory and called up as required. You can store 999 reference weights in a fail-safe memory. Each memory location has: > three-digit sequential number > Designation (alphanumeric max. 18 digits) >...
Direct call via memory location number. Enter the memory location number using the numeric keypad. Direct call via barcode or EDP (See chapter 7.10 Bar code management for information or ask your Soehnle Industrial Solutions service partner). Note If the function read-only memory call via barcode is activated and one is called that is not contained in the fixed memory, the error message "Error 83 - Fixed value not available"...
6.5.14 Transfer of current values to the fixed memory To store fixed tare values directly in the fixed memory, first store the tare value or enter the tare manually. Press the F5 key to confirm. Fix, the value can be stored directly in the Fix memory. The input window for the fixed tare values appears.
6.5.16 Settings in the Setting Mode You can use the parameters in the setting mode to adapt the weighing and taring application program to your individual requirements. To enter the setting mode, press F6 Setting mode in the Setting functions level and then press F6 Apply to call up the program settings.
Page 68
Weighing and taring Display Information field Selection of which data is to be displayed in the information field > Factory default: Change between the individual functions used, e.g. switching functions outputs and inputs and organizational data memory. > IO control: Continuous display of output and input states. >...
6.6 Totalising and picking Totalising means that the individual positions are weighed one after the other, the weight is added to the totalising memory and the scale is unloaded again after each individual weighing. Picking means that the weighed positions remain on the scale. The weight is added to the total memory and the scale returns to zero without being unloaded.
6.6.2 Display view After the first totalising process, the display changes from the basic view to the totalising view, provided that the setting Display information field is set to factory setting. Hint: To additionally display the 10 times higher resolution x10 or the tare weight, for example, you can activate an additional auxiliary display by pressing the F3 key.
6.6.3 Display information field In the Setting Mode / Program Settings / Totalising + Picking / Display Information field can be used to set which data is displayed in the information field. The following options are available: > Factory default: Switch between the individual functions used, e.g. sum memory, organizational data memory and switching functions for outputs and inputs.
6.6.5 Operating the totalising function Use the function change key to call up the totalising functions level. Plus F6 Press the F6 Plus key to transfer the weight on the weighbridge to the totalising memory. The transaction counter and the sequential number are incremented by 1. Before totalling again, the scale must be unloaded to at least the unload factor specified in the setting mode.
Info: Auto totalising can be defined as a print-triggering condition in the same way as manual totalising (see section 8 Printing in the 3025 Operation Manual). With auto batching, a load <5 digits is not automatically accepted; weights < 5 digits can be manually stored in the pick memory using the F6 Plus key.
6.6.10 Assigning and deleting the sequential number Each entry (plus, manual, cancellation) in the total memory is assigned a four-digit sequential number. The sequential number is not reset in the factory setting. Optionally, the consecutive number can be reset in the setting mode: >...
Display Current totals view: Note: Display information field with current totals In the Setting Mode / Program Settings / Totalising + Picking / Display Information field can be set to sum, then the sum is always displayed in the information field. 6.6.14 Display individual items Within a total, you can call up the individual items from the totalising memory by pressing and the F5...
6.6.15 Display options with the Info key Pressing the key gives you the following options: > F1 Tare fix Display of the stored tare fixed values (Tare-Fix memory). > F4 Total Display of the current totals in the totalising memory >...
By pressing the F5 Apply Fix key, the value can be stored directly in the fixed memory. The input window for the fixed tare values appears. To enter the designation of the fixed tare value, press the F6 Edit key. Enter the name using the keypad and confirm your entry by pressing the F6 Enter key.
Page 78
Totalising and picking Auto total/pick Selection off or on On means that only the first position at a time needs to be stored by pressing the Plus key, all other positions are automatically placed on the scale after weighting and stability is stored in the totalising memory. Factory setting: Off Selection of the condition under which the consecutive totalling / picking number Reset consecutive number...
6.7 Counting Counting of parts of the same weight. The individual part weight (reference weight) is determined from a known small quantity of counted parts. This is then used as a divisor to determine the number of pieces from the weight of an unknown quantity of these counted parts.
6.7.2 Display view Counting Hint: To additionally display, for example, the reference quantity, the x10 resolution or the tare weight, you can activate an additional auxiliary display by pressing the F3 key. You can set the value to be displayed in the auxiliary display in the setting mode under Setting mode / Terminal / Display / Auxiliary display.
6.7.3 Display information field In the Setting mode / Program settings / Counting / Display information field you can set which data are displayed in the information field. The following options are available: > Counting: Switching between the individual functions used, counting with the reference weight and weight value in gross / net, e.g.
6.7.5 Counting function The reference weight can be determined using the following function keys: > Function key F1 Ref 10 Reference weight determination from a preassigned reference sample quantity. Place 10 counted pieces on the scale, then press the F1 Ref 10 key. The scale determines the reference weight and displays "10 pieces.
Page 83
a) Call the Ref fix memory to edit the counted items: Press key and then F2 Ref Fix Selection of existing fixed values from the list Scroll with the arrow keys F3 or F4 until the desired memory location is reached, accept the desired fixed value with F6 Edit.
When the fixed value is called up, nominal quantities and tolerances for the piece check are also accepted, if saved with . To delete all stored reference weights, use the "Delete reference weights" function in the setting mode. Hint If the function read-only memory call via barcode is activated and one is called that is not contained in the fixed memory, the error message "Error 83 - Fixed value not available"...
Page 85
Display view After the first totalisation, the display changes from the Count basic view to the Totalise view if the setting "Display information field" is set to Count. Operating the Totalising functions Use the function change key to call up the Totalising functions level. >...
Page 86
Selection options for displaying the total Deleting the totalising memory The following functions can be assigned to the Sum function key in the setting mode: > Display, print and delete totals After pressing the Total function key, the totals are displayed for 3 seconds, and the contents of the memory are printed out and deleted simultaneously.
Page 87
Item counter The item counter only counts the actually totalled transactions. For each new transaction with plus, manual or automatic totalisation, the transaction counter is incremented by 1. The transaction counter runs to 999, followed by an error message. In the case of reversal, the item counter is reduced by 1. The item counter is automatically set to zero when the total is deleted.
Display individual items Within a total, you can call up the individual items from the total memory, by pressing the key and the F5 Plus key one after the other. The item list starts with the most recent entry. Use the arrow keys F3 (older entries) and F4 (newer entries) to cycle through the list. The last 50 entries are available.
Example for reference statistics display: Samples Lightest part 24,448 g Heaviest part 24,750 g Divergence 0,302 g Average 24,589 g Printout of statistics Press the Print button. Store the mean value as the reference weight for counting Press F6 Accept. Exiting the reference statistics without accepting the mean value Press F5 Cancel.
Switching points for piece control If you want to transfer the control values via the interface, e.g. to a control light or to a controller, an I/O card must be installed in the terminal. This must be activated in the setting mode under Program settings / Counting / Default IO configuration with Execute.
6.7.10 Counting accuracy In the setting mode, you can influence the counting accuracy of your scale in the Program settings / Counting. > Standard reference sample quantity The reference sample quantity is set to 10 at the factory. You can set this quantity between 1 and 99. For small parts, it makes sense to use a larger number of parts for reference weight determination.
Page 92
Do not place more parts on the scale than are already on it. The reason for this is that a reference weight calculation may only be performed if the quantity on the scale is exactly correct. It would make no sense to calculate the reference weight from 10 parts, add 1,000 parts and then optimize with 1,010 parts.
6.7.11 Display options with the Info key Pressing the button gives you the following options: > F1 Tare fix Display of stored tare fixed values (Tare-Fix memory). > F2 Ref fix Display of stored reference weights (Ref-Fix memory) > F3 Check pieces Display of the current setpoint with tolerance >...
Enter the name of the reference weight by pressing the F6 "Edit" key. Enter the name using the keypad and confirm your entry by pressing the F6 "Apply" key. Enter the article number in the same way. 6.7.13 Organizational data (Identifier) Detailed information on the organizational data can be found in chapter 5.11 on page 46.
6.7.14 Settings in the setting mode You can use the parameters in the setting mode to adapt the Counting application program to your individual requirements. To call up the setting mode, press the F6 Setting mode key in the Setting functions level and call up the program settings there with F6 Apply.
Page 96
Counting +/- tolerance separated Selection off or on Selection on causes that + and - tolerance for the piece control can be defined and queried separately. Selection Off causes the tolerance to be queried only once and is therefore the same for + and -.
Page 97
Counting Total key function Selection of which functions are to be linked to the sum key Display totals only After pressing the function key Total, the total is displayed. The user can then decide whether the total memory is to be deleted or not. Press the F2 key to clear the total memory.
Page 98
Counting Default IO configuration Selection do not execute or execute Execute activates the switching functions for an external control light or controller. To control the switching operations, an I/O card must be installed. Start mode Selection off or on. When the start mode is activated, the scale freezes the value and thus the signal once the activated switching points have been reached.
Page 99
6.8 Checking Check whether the sample corresponds to a target weight within a specified tolerance range. This function can also be used to weigh the sample to a specified target value with an upper and lower tolerance limit. Tolerance deviations from the target value can be determined in the adjusted unit or in %. In the program settings for Checking, you can select setpoint comparison with the current measured value (factory setting) or totalising memory.
Page 100
6.8.2 Display view Hint: For an additional display, e.g. of the setpoint or the difference to the setpoint, you can activate an additional auxiliary display by pressing the F3 key. The value to be displayed in the auxiliary display is set in setting mode under Setting mode / Terminal / Display / Auxiliary display.
6.8.3 Display information field In the Setting mode /Program settings / Checking / Display information field you can set which data are displayed in the information field. The following options are available: > Factory default: Switch between the individual functions used, e.g. control light, totalising memory, organizational data memory.
Page 102
Control status Control light Special characters Below tolerance yellow Within tolerance green Setpoint reached exactly green Above tolerance In the auxiliary display with the control light setting, no symbols are displayed, only the colours in the display field of the auxiliary display. If a memory location has been activated from the Contr.-Fix memory that contains a target and tolerance value, the traffic light / symbol control functions are automatically activated according to the settings.
Page 103
Editing an existing fixed value from the list Edit the selected fixed value via F6 Edit. Parameter Name is highlighted, call via F6 Edit, input via keyboard. Accept the entry with F6 (abort without accepting the entry with F5 Abort). Further parameters are edited in the same way.
Page 104
Switching points during control If you want to transfer the control values via the interface, e.g. to a control light or to a controller, then an I/O card must be installed in the device. This must be activated in the setting mode under Program settings / Checking / Default IO configuration with Execute.
Page 105
Operating the totalising functions Use the function change key to call up the Totalising functions level. Plus F6 Press the F6 Plus key to transfer the weight on the weighbridge to the totalising memory. The transaction counter and the sequential number are incremented by 1. Before totalling again, the scale must be unloaded to at least the unload factor specified in the setting mode.
Page 106
Info: Auto totalising can be defined as a print-triggering condition in the same way as manual totalising (see section 8 Printing in the 3025 Operation Manual). With auto picking, a load <5 digits is not automatically accepted; weights < 5 digits can be manually stored in the pick memory using the F6 Plus key.
Page 107
Display view current totals: Note: Display information field with current totals In the Setting mode / Program settings / Checking / Display information field can be set to total, then the total is always displayed in the information field. Display individual items Within a total, you can call up the individual items from the total memory, by pressing the key and the key F5 Plus one after the other.
Page 108
Totalising inspection goods out of tolerance It is possible to prevent the totalising of control goods that lie outside a specified tolerance range. If the weight value is outside the tolerance range, press the F6 Plus key and the error message Error 27 "Out of tolerance"...
Page 109
The input window for the setpoint appears. The name of the setpoint is entered by pressing the F6 Edit key. Enter the name using the keypad and confirm your entry by pressing the F6 Accept key. The article number is entered in the same way. 6.8.9 Organizational data (Identifier) Detailed information on the organizational data can be found in chapter 5.11 on page 46.
Page 110
Checking Total outside tolerance Selection within or our of Here you define whether summations in the totalising memory may only be made for values within the tolerance or also for values outside the tolerance. Factory setting: out of Selection of Totalising or Picking for the following settings. Total or picking Totalising means that the individual items are removed from the scale after they have been stored in the totalising memory.
Page 111
Checking Fixed values checking Processing of fixed values Controlling Selection of existing fixed values from the list Scroll with the arrow keys F3 and F4. Call the desired fixed value with F6 Edit. Edit an existing fixed value from the list. Press F6 to call up editing of the selected fixed value.
Page 112
Checking Display information field Selection of which data is to be displayed in the information field > Factory default: Switch between the individual functions used, e.g. control light, totalising memory, organizational data memory. > Control display: Continuous display of the traffic light function. >...
> Communication (Alibi memory, Interfaces, Print templates, Barcode, Batch processing, IO control) > Print > Service (Error messages, Software updates) > and much more can be found under: https://www.soehnle-professional.com/en/site/documents Please go to the customer center on our website www.soehnle-professional.com and select the terminal 3025 under Downloads/Industrial scales/Terminals.