Heaters; Air Bath Oven Heaters - Siemens Maxum II PD PA AP Service Manual

Valves and oven components
Table of Contents

Advertisement

Heaters

5.1

Air Bath Oven Heaters

The air bath oven can be setup for a High Power Configuration or Medium Power Configuration.
This table describes the differences between the two configurations. The high power heater
has a red insulation cover and the medium power has a gray insulation cover.
Two air bath ovens; either
● Dual isothermal air bath
● Programmed Temperature
Single isothermal air bath oven operat‐
ing above 70°C control temperature
Single isothermal air bath oven operat‐
ing at 70°C or below control temperature
Airless ovens: either
● Single isothermal airless
● Dual isothermal airless
Heated sample injection valves, heated
Flame Ionization and Flame Photomet‐
ric Detector Modules
The PECM provides the voltage to the two pairs of 1400-Watt or 650-Watt AC Air Bath heater
power switching circuits located on the rear wall of the EC. Connection to the heaters is made
through a special cable. Each circuit consists of two series connected solid-state relays.
Maxum II Valves and Oven Components
Service Manual, 10/2018, A5E42019844001
Configuration
HIGH POWER CONFIGURATIONS
MEDIUM POWER CONFIGRUATIONS
Heater and Temperature Control Hardware
High power heater control relay assembly and two 1400 watt
air bath heater assemblies
High power heater control relay assembly and 1400 watt air
bath heater assembly
Medium power heater control relay board and 650 watt air
bath heater assembly
Medium power heater control relay board and standard heat‐
ers for airless ovens.
Control relays on Power Entry and Control Module (PECM)
5